Place the compressor units on the vibration
reducers with the motor terminal box facing
towards the control.
❯
Attach the compressor unit with the lock wash-
ers and nuts.
4 x
❯
Connect the compressor unit with the mem-
brane drying unit via the pressure hose.
Warning – risk of dangerous electric volt-
ages
❯
The mains plug must not be plugged in. If it is
plugged in, unplug it.
❯
Unscrew the fastening screws of the cover for
the control.
NOTICE
The operating panel cable is very
short and can damage the PCB when
the cover is removed.
❯
Carefully remove the cover of the con-
trol.
❯
Unplug the operating panel cable.
Assembly
9000-610-72/30 1907V007
17