background image

5

AC & DC VOLTAGE MEASUREMENT

• 

Move the rotary dial to 

     

, press the “FUNC.” button and select DC or AC voltage 

measurement.

• 

Connect the probe to be measured to the circuit or power supply in parallel and 

measure the voltage.

• 

Read the measurement on the screen. When measuring DC voltage, the display 

also shows the voltage polarity of the red probe test point.

       Warning

• 

Do not input voltage above 600V as there is a risk of damaging the instrument.

• 

After completing all measurements, always disconnect the probe and the 

measured circuit.

AC & DC CURRENT MEASUREMENT

• 

Move the rotary dial to the  

  position, press the “FUNC.” button and select DC 

or AC current measurement.

• 

Cut off the power supply to the measured circuit and discharge all of the high 

voltage capacitors on the measured circuit.

• 

Disconnect the circuit to be tested and connect the instrument to the circuit to be 

measured in series.

• 

Connect to the power supply of the circuit and then read the measurement on the 

screen. When measuring DC current, the display also shows the voltage polarity of 

the red probe test point.

Note: If “OL” is displayed it means that the instrument exceeds the measurement range 

of the instrument.

• 

Disconnect the circuit to be tested, remove the probe of the instrument and restore 

the circuit.

       

Warning

• 

Do not input current above 200mA, otherwise the fuse in the instrument may be 

burned.

• 

Disconnect the power supply of the measured circuit firstly and fully discharge all 

high voltage capacitors, then connect the instrument to the circuit in series.

• 

After completing all measurements, always disconnect the probe and the 

measured circuit.

 

RESISTANCE MEASUREMENT

• 

Move the rotary dial to the       position. Note: If it’s not in the resistance 

measurement function, press “FUNC.” to switch to resistance measurement 

function.

• 

Connect the probe to the circuit to be measured in parallel and measure the 

resistance.

• 

Read the measurement result of the screen.

       

Warning

• 

Firstly, disconnect the power supply of the circuit to be measured and fully 

discharge all high voltage capacitors.

• 

After completing all measurements, always disconnect the probe and the 

measured circuit.

 

Содержание D03127

Страница 1: ...1 MODEL D03127 Digital Multimeter Pen Type...

Страница 2: ...rview 4 Measurement Operation 5 AC DC Voltage Measurement 5 AC DC Current Measurement 5 Resistance Measurement 6 Diode Measurement 6 Connectivity Test 6 Non contact Voltage Detection 6 General Technic...

Страница 3: ...instrument around explosive gas steam or in a wet environment When measuring connect the zero ground line first then connect the live wire When disconnecting ensure you disconnect the live wire first...

Страница 4: ...unit is turned the default mode is automatic range on or the function is switched the instrument displays AUTO symbol The operations of entering or quitting manual range are as follows In automatic r...

Страница 5: ...screen When measuring DC current the display also shows the voltage polarity of the red probe test point Note If OL is displayed it means that the instrument exceeds the measurement range of the instr...

Страница 6: ...50 the buzzer will sound Warning Firstly disconnect the power supply of the circuit to be measured and fully discharge all high voltage capacitors After completing all measurements always disconnect t...

Страница 7: ...humidity is no more than 80 DC Voltage Range Resolution Accuracy 200mV 0 1mV 0 5 reading 3 2V 0 001V 20V 0 01V 200V 0 1V 600V 1V 0 8 reading 5 Input impedance 10M Maximum input voltage DC or AC 600V...

Страница 8: ...ore than about 50 Testing current about 1mA open circuit voltage about 2 8V Input protection Maximum DC or AC 600V RMS REPLACING THE BATTERIES Turn off the power supply and disconnect the probe from t...

Отзывы: