11
V7501
2
2
4
4
6
6
3
V7502
2
2
4B
4A
6
4B
4A
3A
3B
6
2� Connect the cold water supply to the water tank using the blue, semi-transparent (1/4” OD x 1/8” ID)
polyurethane tubing to the system� Continue the tubing from the outlet of the VistaCheck to the
top stainless hose barb on top of the brass water valve on the thermal regulator� Optional: warm the
end of the tubing with hot water before pushing it over the barb and make sure it is pushed down
completely over the stem of the barb and is flush with the hex stainless base of the barb�
3� Connect the high-temp Teflon 1/4” OD tubing from the autoclave to the milky white Kynar fitting on
top of the white manifold� Make sure the tubing does not kink when moving the autoclave or pulling
the tubing through the cabinet partition holes�
4� Connect the black 1/4” OD LLDPE condensate line tubing to the elbow fitting on the top of the
manifold, then connect the other end to one of the male adapter fittings on the drain adapter�
5� Cut the black condensate line (from step 4) as close to the drain adapter assembly as possible and
install the in-line thermal sensor� Make certain the flow direction on the in-line sensor is oriented
towards the drain�
6� Connect one end of the 1/2” ID clear PVC tubing to the elbow hose barb on the tank and the other end
to the straight 1/2” hose barb on the drain line adapter� Make sure the tubing is firmly engaged over the
barbs� If in doubt, use appropriate size hose clamps�
2
3
4
5
6
5
5B
5A
TO AUTOCLAVE
TO AUTOCLAVES
MANIFOLD
MANIFOLD
MANIFOLD