Trouble Shooting
Error Codes:
Under the staed conditions, Error Codes are shown on the display:
01 = Action Not Allowed at This Time
02 = Parameter Exceeds Limit of Adjustment
03 = A Timeout Occured during Adjustment
05 = Attempt To Change Units While Pascal Lock Engaged
21 = CDG Zero Adjustment Not Allowed at This Pressure
22 = CDG Atmosphere Adjust Not Allowed at This Pressure
Figure 18: Error Codes
Self Test:
If the self-test fails, run the self-test again by turning the power off and then on again.
The folowing sequnce should be executed:
• “beeper”
• indicators for
TORR
,
MTORR
,
MBAR
,
Μ
BAR
,
PASCAL
,
KPASCAL
,
GAGE
1,
GAGE
2
• 10 LED indicators for set points and other functions
• all four digits will light, including decimal points
• display shows the model number of the instrument,
908A
• display shows software version, e.g.
1.10
If it fails again, call Duniway Stockroom.
Fuses:
If fuses burn out, check to see that the proper voltage has been supplied to the power input
module. If fuses burn out repeatedly call Duniway Stockroom.
Display Shows -LO, OFF or HI:
If the digital display consistently shows
-LO
,
OFF
or
HI
, it
may be that one of the internal power supply protection devices has removed power to the CDG.
You may check this by measuring the voltage at the connector or cable for the unaffected gauge.
Since power for both gauges use the same protection device, either connector will show the power
supply voltages. Normal range for the voltages are 14.5 to 15.5 volts for both +15 volts and -
15 volts. +15 may be measured on the red wire; -15 is on the green wire; power return is on the
black wire. If the power supply protection has shut the power off, you will typically measure less
than 4 volts on the affected supply.
If you verify that either power supply is shut off, remove power from the CDG for a few minutes
to allow the protection device to reset itself. The protection device does not need to be replaced;
it is a reusable thermal fuse.
You may wish to determine the cause for the loss of power supply voltage before applying power
again. The 908A will protect itself if it finds excessive power draw again.
It is normal for the 908A to feel warm to touch along the left side of the case. This is especially
true when operating heated CDGs because of the greater power they require.