D U N I W A Y S T O C K R O O M C O R P .
24 of 29
P. Pascal/kPascal Pressure Units Only Set-Up (-J Mode)
The Terranova 908A has a special -J Mode, where only Pascal/kPascal units are available for dis-
play. In this mode, during the SELECT process, the UNITS will not permit changing to Torr/
mTorr or mBar/microBar. This mode is intended for use in the Japanese market and elsewhere if
choice of units is not allowed.
To enter the -J mode, unplug the 908A from its power source. Press and hold all three
RAISE
,
LOW
-
ER
and SELECT buttons at the same time; while holding the
THREE
buttons depressed, plug the
power cord in. Maintain the depression of three buttons for a few second until the -J mode has
been set up. The digital display will show, in the revision number sequence
N.NNJ,
to confirm
the reset has been entered.
The unit will not return to normal operation for a normal RESET operation. To exit the -J mode,
repeat the process shown above: To exit the -J mode, unplug the 908A from its power source.
Press and hold all three
RAISE
,
LOWER
and SELECT buttons at the same time; while holding the
THREE
buttons depressed, plug the power cord in. Maintain the depression of three buttons for a
few second unitl -J mode has been exited. The digital display will then show, in the revision num-
ber sequence
N.NN,
to confirm the reset has been entered. At this point, the “units” choice can
made using SELECT, RAISE and LOWER buttons
Low Range
High Range
20
mTorr
2.66
Pa 0.000 to 2.660
50
mTorr
6.65
Pa 0.000 to 6.650
100
mTorr
13.3
Pa 0.000 to 9.999 10.00 to 13.30
1
Torr
133
Pa 0.00 to 99.99
100.0 to 133.0
2
Torr
266
Pa 0.00 to 99.99
100.0 to 266.0
5
Torr
665
Pa 0.00 to 99.99
100.0 to 665.0
10
Torr
1330
Pa 0.0 to 999.9
1000 to 1330
20
Torr
2660
Pa 0.0 to 999.9
1000 to 2660
50
Torr
6650
Pa 0.0 to 999.9
1000 to 6650
100
Torr
13.3
kPa 0 to 9999 Pa
10.00 to 13.30 kPa
200
Torr
26.6
kPa 0 to 9999 Pa
10.00 to 26.60 kPa
500
Torr
66.5
kPa 0 to 9999 Pa
10.00 to 66.50 kPa
1000
Torr
133
kPa 0.00 to 99.99
100.0 to 133.0
2000
Torr
266
kPa 0.00 to 99.99
100.0 to 266.0
5000
Torr
665
kPa 0.00 to 99.99
100.0 to 665.0
Terranova 908A-J
CDG Full Scale Range
mTorr/Torr
Pascal/kPascal
Digital Display
Figure 17: Torr-Pascal Relationship Terranova 908J
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: [email protected]