6 -
①
Use this reader on a flat surface avoiding direct light.
②
To prevent electric shock, check if the power supply is connected to the ground.
③
Use supplies only from DSM to prevent product damage.
④
Be careful not to block the air passage on the back of the reader.
⑤
Be careful not to get burnt from the heat produced by the reader due to long-term
use.
⑥
Do not use this reader in a place with shock and vibration.
⑦
To prevent contamination of the reader, wait until all samples are absorbed into
the RDT (Rapid Diagnostic Test) device before you insert the device in the system.
⑧
While you are operating the reader with the object of testing, prevent any vibrations
that affect the system.
⑨
If there is any vibration affecting the system during the test, abort the process and
restart the test from the beginning.
⑩
Do not open the tray during the test. If the device is exposed to the outside during
the test, abort the process and restart the test.
⑪
When the test is completed, exit the program and turn off the system. If you turn
off the system before the program normally ends, it could affect the program and
unexpected errors could be generated.
⑫
Be careful not to contaminate the reader when taking the device out of the reader.
Notice on Copyrights
①
All rights are reserved by DSM.
②
This user manual cannot be copied without permission. If it is used without
authorization, the user has to take legal responsibilities.
CAUTION: If you need an accurate result or have an important test to be done,
wait about 30 minutes after turning the system on and do the
calibration before the test (Refer to 3.2 Instrument Check).