ENGLISH
*Checking point
Check whether the tubing is bent.
Check if there is something blocking the tubing.
Check whether air socket is blocked.
- Please contact your local representative.
2) Code 21 (Low pressure error)
If pressures lower than 30% of the set value are detected more than 3 times during
operation, it will indicate Error code 21.
3) Code 22 (High pressure error)
If pressures higher than 30% of the set value are detected more than 3 times during
operation, it will indicate Error code 22.
4) Code 31 (Vent error)
Error code 31 occurs when the pressure sensor inside the product malfunctions.
*Checking point
Check the connection of the device to the tubings
(tubing).
Check that each sleeve is wrapped properly on the
patient’s leg.
Check whether there is any leakage from a
connected sleeve.
Separate the sleeve and tubing, check air is leaving
from the air socket of the device.
5) Code 32 (Pump error)
During the initial self detection test, if the pump is not operating normally, Error code 32
will be displayed.