Control Panel Functions

Time Adjustment and Timer Operation:

 The run time may be set from 30 seconds to 99 minutes 

and 30 seconds. Press the up and down arrow buttons next to the time display to adjust the run 
time. Adjustments may be made prior to a run. A quick tap will adjust the time by 30 seconds. Hold 
down the button for 1 minute adjustments.

Speed Adjustment:

 The speed may be set from 500 RPM to 3,400 RPM. Press the up and down 

arrow buttons next to the speed display to adjust the RPM. Adjustments may be made prior to a 
run. A quick tap will adjust the speed in increments of 50 RPM. Hold down the buttom to adjust the 
speed in increments of 100 RPM.

Starting and Stopping a Run:

 With the lid switch closed, press the 'START' button to begin a run. 

Press the 'OPEN / STOP' button at any time during a run to terminate it.

Unlocking the Lid:

 The lid is unlocked whenever the red 'UNLOCKED' indicator light is illuminated. 

The lid unlocks automatically for 60 seconds at the end of a run. The lid can also be unlocked for an 
additional 15 seconds by pressing the 'OPEN / STOP' button while the unit is idle. The lid cannot be 
unlocked while the rotor is spinning.

Imbalance Detection Reset:

 This centrifuge is equipped with imbalance detection. If an

imbalance is detected, the centrifuge will terminate the current run and will begin to brake to a stop.
The words 'BALANC' and 'ERROR' will flash on the speed display. Once the rotor has stopped,
open the lid to cancel the error reporting, balance the load and begin a new run. Alternately, the
error reporting can be canceled by pressing the 'OPEN / STOP' button.

9

Содержание Owl Manor HORIZON 24 Flex

Страница 1: ...Laboratory Centrifuge Model HORIZON 24 Flex Operator s Manual P N 03 0 0002 0143 Rev C MODEL HORIZON 24 FLEX...

Страница 2: ...7 Operation 7 Guide to Balanced Centrifuge Loads 8 Control Panel Functions Time Adjustment and Timer Operation 9 Speed Adjustment 9 Starting and Stopping a Run 9 Unlocking the Lid 9 Imbalance Detectio...

Страница 3: ...f tubes Cycle settings can be changed to accommodate custom setting WARRANTY Drucker Diagnostics warranties that this centrifuge is free from defects in workmanship and parts for 2 years See the next...

Страница 4: ...minutes 1 Weight 39 lbs Ambient Temperature During Operation 16 o C 35 o C Maximum Relative Air Humidity 90 Power Watts 280 Voltage Volts 95 to 235 Frequency Hz 50 60 Specifications General Specifica...

Страница 5: ...ol Panel see below Cabinet Up Down RPM Change Buttons Running Indicator Light Latched Indicator Light Unlocked Indicator Light RPM Display Open Stop Button Start Button Up Down Time Change Buttons Tim...

Страница 6: ...atched properly The centrifuge will not run unless the lid is latched properly and the LATCHED indicator light is illuminated 8 Use the up arrow button next to the RPM display to set the speed to full...

Страница 7: ...nd the rotor will decelerate to a complete stop 8 The UNLOCKED indicator light will illuminate the lid light will flash and the locking mechanism will disengage allowing entry into the rotor chamber I...

Страница 8: ...r a device and a counterbalance Please refer to device I F U for proper use of the counterbalance 1 device Load the opposing bucket with a counter balance 2 devices Place the two devices in opposite b...

Страница 9: ...lid switch closed press the START button to begin a run Press the OPEN STOP button at any time during a run to terminate it Unlocking the Lid The lid is unlocked whenever the red UNLOCKED indicator li...

Страница 10: ...ly remove the bucket Properly dispose of the sample and tube fragments and thoroughly clean both the inside and outside of the bucket Insert a new tube cushion if necessary and replace the bucket in t...

Страница 11: ...ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP button to cancel the error and then check...

Страница 12: ...l remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully clockwise to its stop position and the centrifuge will resume operation Emergency Rotor Chamber...

Страница 13: ...nd recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment For more...

Отзывы: