Drucker Diagnostics Owl Manor HORIZON 24 Flex Скачать руководство пользователя страница 10

Care and Preventative Maintenance

With proper care and maintenance your HORIZON centrifuge will provide years of laboratory service. 
For proper care, the following steps should be taken:

1. Provide Adequate Ventilation:

 

For cooling purposes, the HORIZON draws in ambient air through 

the air intake cover on the top of the lid and exhausts this air in the back of the base. The centrifuge 
should be placed on a hard smooth surface for good air circulation.

2. Always Spin Balanced Loads:

 This centrifuge is equipped with imbalance detection. However, it is 

still possible to run loads that are minimally imbalanced.  Refer to page 8 for additional information on 
balancing the load.

3. Keep the Buckets Clean:

 

NOTE: Always follow the safety guidelines of your laboratory to properly 

clean up and/or dispose of materials in the event that a substance known to be potentially toxic, 
radioactive or contaminated with a pathogenic microorganism is spilt in or on the centrifuge. Small 
glass fragments left in the bucket after a tube breakage may adhere to the next tube inserted in that 
bucket. When this tube is handled, these fragments may puncture protective gloves and lacerate the 
operator’s fingers or hand. Remaining fragments may provide stress points on subsequent tubes and 
result in additional breakage. If a tube breakage occurs, carefully remove the bucket. Properly dispose 
of the sample and tube fragments and thoroughly clean both the inside and outside of the bucket. 
Insert a new tube cushion (if necessary) and replace the bucket in the rotor.

4. Motor and Electrical Maintenance:

 This centrifuge uses a brushless-DC motor. There are no 

brushes to replace and it should not need routine servicing for the life of the centrifuge. The electrical 
components are selected for high reliability and should not need service.

5. Keep the Centrifuge Clean:

 

The cabinet, rotor top and accessories shall be thoroughly cleaned 

using either soap and water, or a mild bleach solution. It is critical that the buckets are dried completely 
inside before any product tubes are loaded. Left over chemical residue can damage or weaken the 
tubes. The use of Fully/Partially Halogenated Hydrocarbons, Ketones, Esters and all other chemicals 
not prescribed by the manufacturer may cause damage to the rotor and buckets and shall not be used.

If it is necessary to remove the rotor for additional cleaning it is required that a qualified technician remove 
the outside housing and rotor assembly. Contact your authorized dealer for additional information.

Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in water or other cleaning 
solutions as this will cause damage and void your warranty!

6. Bucket  Replacement:

 It is recommended that the buckets be replaced after 24 months of use. 

Inspect buckets regularly for cracks. If cracks are discovered, replace immediately.

10

Содержание Owl Manor HORIZON 24 Flex

Страница 1: ...Laboratory Centrifuge Model HORIZON 24 Flex Operator s Manual P N 03 0 0002 0143 Rev C MODEL HORIZON 24 FLEX...

Страница 2: ...7 Operation 7 Guide to Balanced Centrifuge Loads 8 Control Panel Functions Time Adjustment and Timer Operation 9 Speed Adjustment 9 Starting and Stopping a Run 9 Unlocking the Lid 9 Imbalance Detectio...

Страница 3: ...f tubes Cycle settings can be changed to accommodate custom setting WARRANTY Drucker Diagnostics warranties that this centrifuge is free from defects in workmanship and parts for 2 years See the next...

Страница 4: ...minutes 1 Weight 39 lbs Ambient Temperature During Operation 16 o C 35 o C Maximum Relative Air Humidity 90 Power Watts 280 Voltage Volts 95 to 235 Frequency Hz 50 60 Specifications General Specifica...

Страница 5: ...ol Panel see below Cabinet Up Down RPM Change Buttons Running Indicator Light Latched Indicator Light Unlocked Indicator Light RPM Display Open Stop Button Start Button Up Down Time Change Buttons Tim...

Страница 6: ...atched properly The centrifuge will not run unless the lid is latched properly and the LATCHED indicator light is illuminated 8 Use the up arrow button next to the RPM display to set the speed to full...

Страница 7: ...nd the rotor will decelerate to a complete stop 8 The UNLOCKED indicator light will illuminate the lid light will flash and the locking mechanism will disengage allowing entry into the rotor chamber I...

Страница 8: ...r a device and a counterbalance Please refer to device I F U for proper use of the counterbalance 1 device Load the opposing bucket with a counter balance 2 devices Place the two devices in opposite b...

Страница 9: ...lid switch closed press the START button to begin a run Press the OPEN STOP button at any time during a run to terminate it Unlocking the Lid The lid is unlocked whenever the red UNLOCKED indicator li...

Страница 10: ...ly remove the bucket Properly dispose of the sample and tube fragments and thoroughly clean both the inside and outside of the bucket Insert a new tube cushion if necessary and replace the bucket in t...

Страница 11: ...ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP button to cancel the error and then check...

Страница 12: ...l remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully clockwise to its stop position and the centrifuge will resume operation Emergency Rotor Chamber...

Страница 13: ...nd recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment For more...

Отзывы: