Drucker Diagnostics Owl Manor HORIZON 24 Flex Скачать руководство пользователя страница 6

6

If any problems are found during the initial setup procedure, refer to the troubleshooting section. 

1.   Plug the female end of the line cord into the back of the centrifuge. Plug the male end into an 

approved electrical outlet. For electrical safety, the unit must always be properly grounded.

2.  Flip the power switch located on the back of the centrifuge to the ON position.
3.   For operator safety, the locking system is always active; requiring power and direction from the 

user to disengage it (the lid also automatically unlocks at the end of a run when it is safe to do 
so). To unlock the lid (in order to access the rotor chamber) press the ‘OPEN / STOP’ button 
on the control panel. The ‘UNLOCKED’ indicator light should illuminate. If it does not, refer to 
the section on troubleshooting. The lid will remain unlocked for 15 seconds after pressing the 
‘OPEN / STOP’ button. 

4.  Turn the latch counterclockwise and open the lid.  
5.  Spin the rotor by hand; check for free and level rotation. If the rotor does not spin freely, refer 

to the section on troubleshooting.

6.  Place the buckets inside the rotor and verify that they are seated properly.
7.  Close the lid. Rotate the lid knob clockwise to its complete stop position. The ‘LATCHED’ indicator 

light should be illuminated. If it is not, make sure that the lid is latched properly. The centrifuge 
will not run unless the lid is latched properly and the ‘LATCHED’ indicator light is illuminated.

8.  Use the up arrow button next to the RPM display to set the speed to full. The top speed for the 

horizontal rotor is 3500 RPM (3200 RPM for fixed angle rotors). 

9.   Initiate a test run by pressing the ‘START’ button.  

(Continued on next page)

1. Unpack the centrifuge and verify that all of the supplied equipment is present.
2. Choose a setup location which meets the following criteria:

a) A bench top clearance height of 21” is required in order to open the lid.
b) The clearance envelope is the space around the centrifuge which is required for safety. Choose a 

setup location which will allow for a clearance envelope of at least 30” x 30”, (with the centrifuge 
at the center). No person or hazardous material shall be permitted in the clearance envelope 
during operation.The operator time within the envelope shall be limited to the time necessary 
for loading, unloading and centrifuge operation only. 

c) Proper ventilation is necessary to prevent the overheating of samples as well as premature 

failure of the centrifuge. Choose an area which will allow unencumbered air flow.

d) The centrifuge is designed to rest on its four rubber feet. No adjustment is necessary for leveling 

the centrifuge, however, the surface should be flat and level.

e) Be sure the outlet is always within reach as the line cord is the means of emergency disconnection.

Initial Setup Procedure

Setup Location

Содержание Owl Manor HORIZON 24 Flex

Страница 1: ...Laboratory Centrifuge Model HORIZON 24 Flex Operator s Manual P N 03 0 0002 0143 Rev C MODEL HORIZON 24 FLEX...

Страница 2: ...7 Operation 7 Guide to Balanced Centrifuge Loads 8 Control Panel Functions Time Adjustment and Timer Operation 9 Speed Adjustment 9 Starting and Stopping a Run 9 Unlocking the Lid 9 Imbalance Detectio...

Страница 3: ...f tubes Cycle settings can be changed to accommodate custom setting WARRANTY Drucker Diagnostics warranties that this centrifuge is free from defects in workmanship and parts for 2 years See the next...

Страница 4: ...minutes 1 Weight 39 lbs Ambient Temperature During Operation 16 o C 35 o C Maximum Relative Air Humidity 90 Power Watts 280 Voltage Volts 95 to 235 Frequency Hz 50 60 Specifications General Specifica...

Страница 5: ...ol Panel see below Cabinet Up Down RPM Change Buttons Running Indicator Light Latched Indicator Light Unlocked Indicator Light RPM Display Open Stop Button Start Button Up Down Time Change Buttons Tim...

Страница 6: ...atched properly The centrifuge will not run unless the lid is latched properly and the LATCHED indicator light is illuminated 8 Use the up arrow button next to the RPM display to set the speed to full...

Страница 7: ...nd the rotor will decelerate to a complete stop 8 The UNLOCKED indicator light will illuminate the lid light will flash and the locking mechanism will disengage allowing entry into the rotor chamber I...

Страница 8: ...r a device and a counterbalance Please refer to device I F U for proper use of the counterbalance 1 device Load the opposing bucket with a counter balance 2 devices Place the two devices in opposite b...

Страница 9: ...lid switch closed press the START button to begin a run Press the OPEN STOP button at any time during a run to terminate it Unlocking the Lid The lid is unlocked whenever the red UNLOCKED indicator li...

Страница 10: ...ly remove the bucket Properly dispose of the sample and tube fragments and thoroughly clean both the inside and outside of the bucket Insert a new tube cushion if necessary and replace the bucket in t...

Страница 11: ...ERROR SPEED The centrifuge cannot reach full speed due to a problem with the rotor an inadequate power supply or other electrical problems Press the OPEN STOP button to cancel the error and then check...

Страница 12: ...l remain locked If the knob is accidentally moved and this situation should occur rotate the knob fully clockwise to its stop position and the centrifuge will resume operation Emergency Rotor Chamber...

Страница 13: ...nd recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment For more...