Copyright 2022 Baker Hughes Company.
8 | UNIK 5*00 Series–English
3.4 Power Requirements
The sensor should be connected to a compliant stable power supply. The power supply must be
energy-limited to a maximum of 4.2 A. The power supply requirements for the sensor are shown
in Table
Table
3: Supply Voltage
Product
Output Type
Supply Voltage
5000
PMP (Basic)
7 to 32
Vdc (12 to 32
Vdc for 0 to 10
V Output)
PMP (Ratiometric)
5 ± 0.5
Vdc
PMP (Basic Configurable)
(Maximum 1
V) (7
V Minimum) to 32
V
PMP (Configurable 3 & 4 wire)
7 to 36
Vdc
PTX
7 to 32
Vdc
PDCR (Passive)
2.5 to 12
Vdc
PDCR (Linearized)
7 to 12
Vdc
5600/5700
PTX
7 to 32
Vdc
5800/5900
PMP (Basic)
7 to 32
Vdc (12 to 32
Vdc for 0 to 10
V Output)
PMP (Basic Configurable)
(Maximum 1
V) to 32
Vdc
PTX
7 to 32
Vdc
PDCR (Passive)
2.5 to 12
Vdc
PDCR (Linearized)
7 to 12
Vdc
Table
4: Supply Consumption
Product
Output Type
Current Consumption
5000
PMP (Basic)
<3
mA
PMP (Ratiometric)
<3
mA
PMP (Basic Configurable)
<3
mA
PMP (Configurable 3 & 4 wire)
<20
mA at 7
Vdc decreasing to <5
mA at 32
Vdc
PTX
4 to 20
mA (Limited to 30
mA maximum)
PDCR (Passive)
<2
mA at 10
Vdc
PDCR (Linearized)
<3
mA
5600/5700
PTX
4 to 20
mA (Limited to 30
mA maximum)
5800/5900
PMP (Basic)
<3
mA
PMP (Basic Configurable)
<3
mA
PTX
4 to 20
mA (Limited to 30
mA maximum)
PDCR (Passive)
<2
mA at 10
Vdc
PDCR (Linearized)
<3
mA
Содержание UNIK 5 00 Series