Charger plugged in
◆
Unplug charger
Control box completely dead
◆
Battery completely flat
◆
Charge battery
on battery and no relay
◆
Battery defective
◆
Replace battery
clicking is heard.
Control box okay apart from
◆
handset defective
◆
Send handset for repair
One direction on one channel
◆
Control box defective
◆
Send control box for repair
Warning and Caution!
1. Patient lifter is a Transfer Device that allows patient attendant to transfer patients from bed
to wheelchair or commode…etc. It however should not be used to transport patients.
2. When lifting patient, make sure the base legs are in the most widely opened position and
the rear caster brakes are engaged. Otherwise, the lifter may tilt over.
3. When battery power is 50% low, the controller will give “beep” warning. At this moment,
you should charge the battery immediately. Please note that a lifted patient cannot be
lowered down when battery is out of power.
4. Push the “Emergency” red button if the control unit system is out of control.
5. “QUICK RELEASE” system (push/pull) enables stepless retraction of the actuator under
load (Mechanical emergency lowering). Refer to Picture A.
Picture A:
QUICK RELEASE
6. How to Charge: (please refer to Picture B )
a. With Jumbo Linak set (battery box, control box & actuator) assembled, insert the power
cord plug in control box and the other end to a wall outlet.
b. Now, it is charging, the “Charge” light is on.
c. When the charging action is finished, the “Charge” light will be off automatically. (Usually,
it is need to charge 4 hours from the 50% capacity to 100% capacity)
d. Remove the power supply cord.
e. Now, you can operate again.