How to Operate the Stand-Aid Lifter: (Weight Lifting Capacity: 180 kgs) 

       

 

Important: Prior to first use, please make sure that the battery is being charged 24 

hours in order to reach proper function and prolong the lifetime of battery. 

 

1. Spread the base legs to the widest position before lifting. 
2. Keep the patient centered between the base legs and have the patient face the attendant. 
3. Be extremely cautious and use restraint straps for spastic or severely handicapped 

patients. 

4. How to use Standard or Commode Slings: 

a. For smooth and easy lifting, have the lift, chains, sling, commode or wheelchair in ready 

position. 

b. With the patient in center of bed, roll on side away from the attendant. 
c. Roll the patient to side toward the attendant and center the patient on sling.    With base 

of lifter under bed, press the “

” button on the control handset to lower the boom (#20).        

d. Hook the hanging strips of sling with the hanging bar carefully.    If a chain is used, make 

sure the “S” hooks are away from the patient.   

e. Lock rear casters, lift the patient by pressing the “

” button on the control handset.   

f.  Lift patient until his or her feet will swing easily off the bed keeping patient facing the 

attendant.                                                                   

g. Unlock rear casters and transfer patient to and above commode or wheelchair.    Lock 

brakes of both lifter and commode (or wheelchair). 

h. Press the “

” button on the control handset to gradually lower the patient. 

i.  During descent, assist patient to attain correct sitting posture. 
j.  For transferring to wheelchair, when patient is seated, push down on boom to slacken 

hanging strips of sling.    Patient can remain seated in sling. 

k. For commode use, adjust clothing before moving lift to straddle commode.    Keep 

hanging strips taut and make sure patient is in a comfortable position. 

 

                              Trouble Shooting (Linak LA34 System) 
 

        Normal Symptom                          Possible Cause                                            Action                                             

Power indicator does not light         

 Not connected to mains       

◆ 

Connect to mains 

up.                                                       

 The fuse has blown           

  Replace fuse, if the system is prepared for external 

fuse replacement, or send the system to repair 

                                                             

 Control box defective         

  Send control box to repair                 

Power indicator lights up         

  Actuator plug not pushed into   

  Push actuator plug into   

But actuator does not run.          control properly                  control box properly              

Relays in control box are        

  Actuator defective                         

  Replace actuator                                 

heard clicking.                 

◆ 

Control box defective          

 Replace control box       

              . 

Power indicator lights up        

◆ 

Control box defective          

 Send control box repair             

but actuator does not run.       

◆ 

Handset defective             

send handset repair                                                           

No relay noise is heard                 

◆ 

Handset plug not insert properly 

◆ 

Check all cord connections                                               

from control box.                           

◆ 

Emergency stop engaged       

 Deactivate emergency stop          

Содержание 13246

Страница 1: ...USER MANUAL Item 13246 Stand Aid Lift Drive Medical Design and Manufacturing www drivemedical com ...

Страница 2: ... qualified professional z The lift must be used by a caregiver with proper training to work with the person to be transferred z Only trained and qualified caregivers should transfer a person Do not attempt to use the lift if you have not been properly trained to do so z Always be prepared before attempting to transfer a person z Do not use a sling that is not recommended for the lift z Do not stor...

Страница 3: ...ts from the carton and place on the floor taking care to protect the finish from damage 2 Place the chassis in a clear space and apply the rear bakes Attach the foot plate to the base Drive Medical Design and Manufacturing www drivemedical com ...

Страница 4: ...3 Fit the mast and boom assembly into the base socket Tighten the mast assembly with the drive screw Drive Medical Design and Manufacturing www drivemedical com ...

Страница 5: ...Tighten up screw Assemble actuator bottom and top Insert the pin chain through actuator top Drive Medical Design and Manufacturing www drivemedical com ...

Страница 6: ...Fit control box onto bracket on mast Tight up control box Fit battery onto control box Drive Medical Design and Manufacturing www drivemedical com ...

Страница 7: ...trol box bottom at left side Plug actuator in control box bottom at middle Fit the adjusting handle REMARKS Please always check the mast is fully locked into position Drive Medical Design and Manufacturing www drivemedical com ...

Страница 8: ...endant g Unlock rear casters and transfer patient to and above commode or wheelchair Lock brakes of both lifter and commode or wheelchair h Press the button on the control handset to gradually lower the patient i During descent assist patient to attain correct sitting posture j For transferring to wheelchair when patient is seated push down on boom to slacken hanging strips of sling Patient can re...

Страница 9: ...e controller will give beep warning At this moment you should charge the battery immediately Please note that a lifted patient cannot be lowered down when battery is out of power 4 Push the Emergency red button if the control unit system is out of control 5 QUICK RELEASE system push pull enables stepless retraction of the actuator under load Mechanical emergency lowering Refer to Picture A Picture...

Страница 10: ...e sure base adjustment mechanism is operating easily e Double check to ensure the U bracket at the end of the boom and the S hook on the hanging bar are secured tightly f Work the Linak Actuator to ensure that the battery still has enough power When battery is 50 below full capacity it will give beep warning and need to re charge immediately g Before initial use place a drop of oil on the followin...

Страница 11: ...Drive Medical Design and Manufacturing www drivemedical com Drive Medical Design Manufacturing 99 Seaview Boulevard Port Washington NY 11050 Phone 516 998 4600 Fax 516 998 4601 www drivemedical com ...

Отзывы: