USER AND MAINTENANCE GUIDE FOR
Samsoft Mini
PATIENT LIFT
6
Photo #°4
- Safety pin
Photo #°5
- Mast safety shaft
Photo #°6
- Position of the safety shaft when the mast
is upright
Photo #°7
- Unfolding the boom
Photo #°8
- Pin of the jack
5.4) Folding
The device should be folded to be transported.
(Photo #°1).
The folding allows to store vertically the device and to transport
him in complete safety. To fold up the device, make the operations
described in the paragraph 4 but in the inverse order :
Press on the remote control to lower the arrow up to the low point.
Remove the jack of the arrow then hang on it on the mast.
Make revolve the arrow towards the back of patient-lift.
Fix the cross-bar to the mast.
Remove the pin (7) and the mast safety shaft (11).
Lower the assembly mast/boom/cross-bar onto the side rails.
CAUTION: pinching risk (9 photo #4)
Reposition the mast safety shaft in its 1st position, and put its
pin back.
The device can be vertically transported or stored (photo #9).
Photo #°9
To facilitate the transport of Samsoft Mini, it is possible to
defuse him in two parts by unscrewing completely the button
(10) photo 4 and by removing the axis of safety (11) photo 5.
Warning Unscrew partially backlash screw (8) photo 4
before separating the mast of the rolling base. After reas-
sembly, verify that all the elements of fi xation and the pin
of safety are fi rmly in place. Squeeze strongly the button (10).
Weight of the rolling basis: 14 kg - weight of the mast, the arrow
and the cross-bar: 21 kg and 17 kg without the pack batteries.
VI - USING
Samsoft Mini
WARNING
Samsoft Mini
should be used indoors exclu-
sively on a fl at and adherent fl oor; it should be handled
by a competent caregiver trained for using this kind of
devices.
7
10
11
12
9
8