8.1 Fault Finding
Listed in the table are a set of faults that may occur within the service life of the
treatment chair. If the fault is not listed please contact your supplier for further
assistance.
If a fault does occur please try the following suggestions, as these may help in
diagnosing
the
fault.
Fault
Possible Cause
Remedy
Electrical function(s)
do not work.
Mains lead not plugged into the
control box or wall.
Check to see if the ‘power on’
light on the control box is lit and
that the mains lead is plugged in
at both ends.
Fuse has blown in the mains plug.
Check to see if the ‘power on’
light on the control box is lit, if not
replace fuse.
Actuator/handset leads not
plugged in.
Check plug connections on the
control box.
Damage to mains cable, actuator
cable or handset cables.
Turn off at the mains and contact
your approved service engineer.
Duty cycle exceeded.
Turn off at the mains and contact
your approved service engineer.
Chair overloaded.
Remove load.
Backrest/leg section
does not raise or
lower.
Lift mechanism/gas strut is faulty.
Contact your approved service
engineer.
Possible obstruction in way of
backrest, leg section or gas strut.
Remove obstruction, operate
leg section/backrest to ensure
function is not compromised.