GENERAL INTRODUCTION

Thank you for selecting the pH6175/6175-3C meter. The

pH6175/6175-3C is a precision tool that measures pH, mV and

temperature. A built-in microprocessor stores, calculates and

compensates for all parameters related to pH determinations

including pH electrode temperature characteristics, electrode slope

deviations and buffer solutions.

This meter has a waterproof IP54 case. The mechanical keys are

highly reliable with tactile and audio feedback. It is powered by six

AAA-size alkaline batteries or with a UL approved AC adapter

(OUTPUT:DC9V). The meter also displays a “BAT” message when

the batteries are in need of replacement. Re-calibration is not

required when power is restored.

The front of the meter has a large LCD with backlight that displays

pH or mV and temperature simultaneously along with user prompts

and mode indicators. The unit prompts the user through calibration

and measurement procedures.

An AUTOLOCK feature for both pH and mV measurements enables

the unit to automatically sense the end point and “LOCK” the display

to indicate the end point value of a measurement. AUTOLOCK and

user prompts help eliminate most errors in determining pH and mV

values,

resulting

in

precise,

repeatable

and

error-free

measurements. The pH6175 can also be used in non-AUTOLOCK

mode.

The model pH6175/6175-3C is available with pH, ORP and ATC

(Automatic Temperature Compensation) probes. Other features

include single or dual or three point calibration (6175-3C include

dual or three point calibration), electrode offset recognition,

electrode slope recognition, electrode efficiency display, built-in

buffer coefficients, automatic or manual temperature compensation,

long battery life, and 50/60Hz AC noise rejection. This meter is

user-friendly for laboratory application.

INITIAL INSPECTION

Carefully unpack the unit and accessories. Inspect for damages

made in shipment. If any damage is found, notify your Drawell

representative immediately. All packing materials should be saved

until satisfactory operation is confirmed.

POWER INPUT

The model pH6175/6175-3C can be powered by an 115V or

Содержание DW-PH6175

Страница 1: ...ure Benchtop Meter Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Страница 2: ...de 8 d Calibration with manual temperature compensation in the pH NON AUTOLOCK mode 9 C pH Measurements 9 a Measurement with an ATC Temp probe in the pH AUTOLOCK Mode 9 b Measurement with manual temperature compensation in the pH AUTOLOCK mode 10 c Measurement with an ATC Temp probe in the pH NON AUTOLOCK mode 10 d Measurement with manual temperature compensation in the pH NON AUTOLOCK mode 10 D T...

Страница 3: ... for both pH and mV measurements enables the unit to automatically sense the end point and LOCK the display to indicate the end point value of a measurement AUTOLOCK and user prompts help eliminate most errors in determining pH and mV values resulting in precise repeatable and error free measurements The pH6175 can also be used in non AUTOLOCK mode The model pH6175 6175 3C is available with pH ORP...

Страница 4: ...ately INSTALLING THE BATTERIES To insert the batteries into the meter follow the procedure outlined below 1 Use two hands to flip the two buckles and battery cover to expose the battery compartment Figure 1 2 Note the polarity and insert the six AAA batteries into the battery compartment correctly 3 Replace the battery cover 4 Figure 1 Battery compartment DISPLAY KEYS FUNCTIONS A Display 1 2 3 4 5...

Страница 5: ...efficiency value is less over than 75 5 ATC MAN ATC indicator will be displayed if a temperature probe is connected otherwise the MAN indicator will be displayed 11 MAIN DISPLAY For pH mV and probe efficiency values 6 CAL This will be displayed when the unit enters into the calibration mode 12 SECONDARY DISPLAY For temperature in display B Keys Stand M ode Clear On Off Press and hold this key for ...

Страница 6: ... the unit when operating in ATC mode M ea Eff Stand Slope M ode Clear M ea Eff Stand Slope M ode Clear Stand Slope The Stand and Slope keys are used for pH calibration of the unit Pressing and holding the Stand key while turning on the power will change the buffer set M ea Eff Stand Slope M ode Clear Mea Eff The key is used to bring the unit out of the AUTO condition when operating in the pH AUTOL...

Страница 7: ...The WAIT icon will flash until the unit detects a stable reading Once the unit calibrates the first point the selected buffer remains lit up while the remaining two buffers start to flash The unit is ready to be sloped at the second buffer Note If the first buffer solution is 7 00 or 6 86 at this moment press the Mode key The unit will exit the calibration mode Single point calibration is complete...

Страница 8: ... pH electrode to the BNC connector of the unit MAN icon will lit up pH icon and AUTO icon will lit up One of the buffer in the pre selected buffer set will start to flash 3 Rinse the pH probes in distilled water and immerse it in the first buffer solution Adjust the temperature reading to that of the first buffer using the up or down keys 0 0 to 60 0 Then press and hold Stand key for 5 seconds to ...

Страница 9: ...ete If the first buffer solution is 4 00 4 01 9 18 or 10 00 at this moment the unit will automatically exit the calibration mode Single point calibration is complete The 6175 3C has not this single point calibration 4 Rinse the pH and ATC Temp probe in distilled water and immerse them in the second buffer solution either 4 00 4 01 or 9 18 10 01 Allow temperature reading to stabilize then press Slo...

Страница 10: ...ining two buffers start to flash The unit is ready to be sloped at the second buffer Note If the first buffer solution is 7 00 or 6 86 at this moment Press the Mode key the unit will exit the calibration mode Single point calibration is complete If the first buffer solution is 4 00 4 01 9 18 or 10 00 at this moment the unit will immediately exit the calibration mode Single point calibration is com...

Страница 11: ...AUTOLOCK mode for measurements b Measurement with manual temperature compensation in the pH AUTOLOCK mode 1 Connect the pH electrode to the BNC connector of the unit The MAN icon will lit up Set unit to display the sample temperature by pressing the up and down keys 0 0 to 100 0 2 Repeat steps 2 5 of Measurement with an ATC Temp probe in the pH AUTOLOCK mode c Measurement with an ATC Temp probe in...

Страница 12: ... AUTO icon lit up 3 Rinse electrode with distilled water and immerse it in sample to be measured 4 Press the Mea key The WAIT icon will start flashing The unit is waiting for a stable reading The display will track the mV value as sensed by the ORP electrode 5 When the WAIT icon disappears the reading is then LOCK and will not respond to further changes from the sample The mV value is the sample r...

Страница 13: ... 9 18 4 01 7 00 10 01 0 4 01 6 98 9 46 4 01 7 11 10 32 5 4 00 6 95 9 39 4 01 7 08 10 25 10 4 00 6 92 9 33 4 00 7 06 10 18 15 4 00 6 90 9 28 4 00 7 03 10 12 20 4 00 6 88 9 23 4 00 7 01 10 06 25 4 00 6 86 9 18 4 01 7 00 10 01 30 4 01 6 85 9 14 4 01 6 98 9 97 35 4 02 6 84 9 10 4 02 6 98 9 93 40 4 03 6 84 9 07 4 03 6 97 9 89 45 4 04 6 83 9 04 4 04 6 97 9 86 50 4 06 6 83 9 02 4 06 6 97 9 83 55 4 07 6 8...

Страница 14: ...the buffer value 2 Buffer 4 00 4 01 9 18 and 10 01 is not correct 3 pH electrode slope is off by more than 30 of ideal slope 1 Allow sufficient time for the electrode and ATC Temp probe to stabilize re press Slope key to continue the calibration procedure 2 Check if the correct buffer is used 3 Replace the buffer and or the pH electrode Press Clear key to recalibrate meter Er3 1 Temperature is out...

Страница 15: ... 0 0 C to 100 0 C pH Buffer Temperature range 0 C to 60 0 C pH Electrode Offset recognition 100 mV at pH 7 00 or 108 3 91 7mV at pH6 86 6175 60 mV at pH 7 00 or 68 3 51 7mV at pH6 86 6175 3C pH Electrode Slope recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input impedance 1012 Ω Temperature sensor Thermistor 10 kΩ at 25 C Power Six AAA Batteries Calibration Back up EEPROM Audio Feedback All Touch K...

Страница 16: ...he product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives before returning items for any reason When applying for authorization have the model and serial number handy including data regarding the reason for return For your protection items must be carefully pa...

Страница 17: ... Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Отзывы: