14
[Note:
If the meter still does not perform normally after the above
measures are taken, call Drawell representative.
]
SPECIFICATIONS
Display
Range
Resolution
Accuracy
pH
0.00 to 14.00 pH
0.01 pH
±0.01 pH
mV(6175-3C)
-1999 to 1999 mV
1 mV
±1 mV
mV(pH6175)
-1999.9 to 1999.9 mV
0.1mV
±0.4/±1 mV
Temperature
0.0 to 100.0 °C
0.1 °C
±0.2°C
pH buffer recognition
pH 7.00, 4.01, 10.01 or pH
6.86, 4.00, 9.18
pH Temperature compensation
AUTO/MAN 0.0°C to 100.0 °C
pH Buffer Temperature range
0°C to 60.0°C
pH Electrode Offset recognition
±100 mV at pH 7.00 or
+108.3/-91.7mV at pH6.86
(
6175
)
±60 mV at pH 7.00 or
+68.3/-51.7mV
at pH6.86
(
6175-3C
)
pH Electrode Slope recognition
±30% at pH 4.00, 4.01, 9.18
and 10.01
Input impedance
>10
12
Ω
Temperature sensor
Thermistor, 10 kΩ at 25°C
Power
Six “AAA” Batteries
Calibration Back-up
EEPROM
Audio Feedback
All Touch Keys
End Point Sensing & Hold
Yes
Display (pH /mV : Temp)
22mm : 14.5mm high LCD
Ambient Temperature Range
0 to 50 °C
Relative Humidity
up to 90%
Case
IP54
Dimensions (W x D x H)
150mm x 203mm x 72mm
Weight
504 grams(Batteries included)