DV-8000 USER’S MANUAL
4
Specifications:
Installation:
1. After carefully unpacking the contents, check the materials with the packing list
(page 2) to ensure that you have received everything in good condition.
2. Place the instrument in a suitable location away from direct sunlight. In order to
have the best performance from your instrument, keep it as far as possible from
any strong magnetic or electrical fields or any electrical device that may generate
high-frequency fields. Set the unit up in an area that is free of dust, corrosive
gases and strong vibrations.
3. Remove any obstructions or materials that could hinder the flow of air under and
around the instrument.
4. Use the appropriate power cord and plug into a grounded outlet.
5. Turn on your DV-8000 model spectrophotometer. Allow it to warm up for at least
20 minutes before taking any readings.
Model
DV-8000
Wavelength Range
325-1020nm
Spectral Bandwidth
4nm
Optical System
Single Beam, Grating 1200lines/mm
Wavelength Accuracy
±2nm
Wavelength Repeatability
1nm
Photometric Accuracy
±0.5%T
Photometric Repeatability
±0.3%T
Photometric Range
-0.097-1.999A 0-125%T
Stray Light
≤
0.2%T @ 340nm
Stability
±0.002A/h @500nm
Display
Graphic LED (4 bits)
Keyboard
Membrane Keypad
Photometric Mode
T, A, C, F
Detector
Silicon Photodiode
Sample Compartment
Standard 10mm path length cuvette
Accommodates 100mm path length cuvette with optional holder
Light Source
Tungsten lamp
Output
RS232 Port
Power Requirement
AC 85-250V
Dimensions (W x D x H)
450 x340 x 180mm
Net Weight
8kg