18
Fig 31
3.3.2
Push the blank cuvette into the Reference Light Path
and Main Light Path, press
【
0Abs/%100T
】
,the
instrument will step to the wavelength and blank. See
Fig 32.
Fig 32
3.3.3
Pull the first sample cuvette of known concentration into
the light path, Press the key
【
START
】
to get values of
standard curve one by one (Fig 33).
3.3.4
Press
【
F4
】
to draw the curve. You can get a different
curve by pressing
【
F1
】
to select a different fitting method.
(See Fig 34-Fig37.) For linear fits, “ r ” represent fitting
coefficient of linear regression .r=1 is best fitting.usually
“ r ” is very close to 1.
Note:If there are few standard samples, it is not suitable for
selecting square fitting, especially cubic fitting, otherwise invalid
fitting result will be obtained.
Fig 33