3. DO NOT OPEN SAMPLE COMPARTMENT LID DURING
BLANKING.
4. The dark current don’t be taken after power on if you bypass the
calibrating system. It is recommended to take the dark current after
warm up.(See page 38).
Fig 9
b
. Take out the sample cuvette, replace the
reference solution
with
sample solution
after flushing the cuvette completely.Put the sample cuvette
into the Sample Light Path, then the result will display on the screen
automatically. However the
【
START
】
must be pressed in other measurements
such as DNA/Protein, Muli WL and Quantitative etc.
B. Take measure in WL Scan
a.
After all scan parameters are entered, put the reference cuvette with
reference solution
into the Reference Light Path and the sample cuvette with
sample solution
into the sample light path,Press
【
START
】
to scan
.
b.
(Recommended) After all scan parameters are entered, put the
reference cuvette with
reference solution
into the Reference Light Path and the
sample cuvette with
reference solution
into the Sample Light Path, Press
【
ZERO
】
to obtain the user baseline
.
Then take out the sample cuvette,
replace the
reference solution
with
sample solution
after flushing the cuvette
completely. Put the sample cuvette into the Sample Light Path. Press
【
START
】
to scan.
※
Set wavelength
(Example: set wavelength in “Basic mode”)
Press
【
SET
λ】
(Fig 10).