V/Hz,
Ω
, Temperature,
and A/Hz Measurements
5.1.3 Alternating Voltage and Frequency Measurement V AC and
Hz with Selectable Low-Pass Filter
∞
In accordance with the voltage or frequency to be measured, turn
the rotary switch to V~ or Hz.
∞
Connect the measurement cables as shown. The “
⊥
” connector
jack should be grounded.
Voltage Measurement
Note!
An intermittent acoustic signal warns the operator if the measured
value exceeds the upper range limit in the 1000 V range.
Make sure that a current measuring range (“A”) has not been activated
when the multimeter is connected for voltage measurement! If the
fuse’s blowing limits are exceeded as a result of operator error, both
the operator and the instrument are in danger!
∞
You can switch back and forth between voltage measurement with
and without low-pass filter.
∞
Press the
FUNC | ENTER
multifunction key repeatedly until unit
of measure V or V/Fil appears at the display.
Frequency Measurement
∞
Apply the measured quantity in the same way as for voltage meas-
urement.
∞
Manually select the measuring range for the voltage amplitude.
When the instrument is switched to frequency measurement, the
previously selected voltage measuring range remains active.
∞
You can switch back and forth between frequency measurement
with and without low-pass filter. Press the
FUNC | ENTER
multi-
function key repeatedly until unit of measure Hz or Hz/Fil appears
at the display. Lowest measurable frequencies and maximum al-
lowable voltages are included in chapter 8, “Technical Data”.
Measurement with Low-Pass Filter
Attention!
!
Be aware of the fact that dangerous voltage spikes are not displayed
during this type of measurement (see also “Voltage Comparator”).
We recommend measuring voltage without the low-pass filter first,
in order to be able to detect any dangerous voltages.
A 1 kHz low-pass filter can be activated if required, in order to filter
out high frequency pulses of greater than 1 kHz, for example when
performing measurements at pulsed motor drives, i.e. undesired volt-
ages of greater than 1 kHz can be suppressed.
The active low-pass filter is indicated by the Fil display. The multime-
ter is automatically switched to manual measuring range selection.
The specified measuring accuracy is not achieved when the filter is
activated and signals are greater than 100 Hz.
Содержание DranTech Xtra
Страница 1: ...DranTech XTRA TRMS Digital Multimeter User s Guide P N M240H 2007 Rev A...
Страница 5: ...DranTech XTRA User s Guide 5 Contact Persons...
Страница 11: ...DranTech XTRA User s Guide 11 Safety Precautions...
Страница 25: ...DranTech XTRA User s Guide 25 Control Functions...
Страница 43: ...DranTech XTRA User s Guide 43 V Hz Temperature and A Hz Measurements...
Страница 49: ...DranTech XTRA User s Guide 49 Device and Measuring Parameters...
Страница 59: ...DranTech XTRA User s Guide 59 Technical Data...