Dräger X-dock 5300/6300/6600
17
6.2
Managing tests
In this menu, the existing tests can be edited, new tests can be
created and the standard test can be defined.
Select
>
Define test
.
The test overview will be displayed.
For more information see chapter 4.8 on page 11.
6.3
Station gas configuration
The individual gas connections can be configured in this menu.
Select
>
Station gas configuration
.
An overview of the test gas connections is displayed.
For more information see chapter 4.7 on page 9.
6.4
System configuration
6.4.1
Setting Date & time
The following settings can be made in this menu:
Time zone
Date format
Date
Time
Location
To set the date and time:
1. Select
> System configuration > Date & time
.
2. Make the desired settings.
3. Confirm settings with
OK
.
6.4.2
Setting the language
1. Select
> System configuration > Language
.
2. Selected the required language.
3. Confirm selection with
OK
.
6.4.3
Network
To set
System address
:
1. Select
> System configuration > Network > System
address
.
2. Select dynamic or static IP address.
3. Where necessary, enter the IP of the system address.
4. Where necessary, enter the Subnet Mask (default:
255.255.255.0).
To set the central database address:
1. Select
> System configuration > Network > DBMS
address
.
2. Enter the server IP #.
6.4.4
Cleaning the database
1. Select
> System configuration > Database > Clean
up
.
2. Select
Start clean up
.
3. Confirm deletion of the database entries with OK.
The progress of the deletion is displayed.
6.4.5
Displaying info screen
1. Select
> System configuration > Info screen
.
The following information is displayed:
Application version
Hardware version
Kernel version
MAC address
IP address
6.4.6
Performing an update
To perform a firmware update, see chapter 9.2 on page 21.
6.4.7
Service interval
1. Select
> System configuration >
Service interval
.
This function displays how many days have passed
since the last service was performed. If the service
interval has been exceeded, an additional check box to
deactivate the self check note for 365 days is displayed.
6.5
Data management
6.5.1
Managing instruments
The individual instrument profiles can be displayed in this
menu.
Select
> Data management > Devices
.
The instrument profiles will be displayed.
For more information see chapter 4.9 on page 13.
6.5.2
Managing user profiles
User profiles can be created, edited or deleted in this menu.
Select
> Data management > Users
.
The user profiles will be displayed.
For more information see chapter 4.6 on page 8.
NOTICE
If the station is connected to a server, only place and
format can be changed. The station synchronises time
and date automatically.
NOTICE
If the IP or DBMS address is changed, the station must
be re-started to make the change active.
i
i
i
i
NOTICE
Database entries, data loggers and certificates are
deleted.
NOTICE
All entries are stored in UTC time (Universal Time
Coordinated).
i