Dräger X-dock 5300/6300/6600
Basics
When using a test gas cylinder from another manufacturer:
1. Create or delete test gas component.
Create a new test gas component with .
Delete the current test gas component with .
2. Select test gas.
3. Enter the test gas concentration.
4. Select test gas units.
5. If required, create additional test gas components.
6. The following information can optionally be given:
Part no. of the test gas cylinder
Batch number of test gas cylinder
Expiry date of the test gas cylinder
7. If required, enter
Further options
.
4.7.1
Further options
1. Enter volume of test gas cylinder.
2. Set type of volume:
Comp.
= cylinder volume
Relax.
= gas volume (cylinder volume x pressure)
3. If required, refresh test gas pressure.
4. If necessary, activate
Hose > 2 m
if the length of the test
gas hose exceeds 2 m.
5. If required, configure other test gas inlets in the same way.
Status of test gas pressure:
Gas cylinder level indicator:
4.7.2
Settings
1. Select
>
Station gas configuration > Settings
.
The OV sensor of the PAC 7000 OV allows to choose between
two different test gases that are also used for calibration and
testing. These test gases are carbon monoxide (CO) and
ethylene oxide (EO).
For the X-am 2000, it is possible to choose between three
different test gases that are also used for calibration and
testing. The three options are methane (CH
4
), propane (C
3
H
8
)
and pentane (C
5
H
12
). The sensor is calibrated with different
sensitivity levels depending on the gas selected. More
information on this can be found in the relevant sensor data
sheets.
NOTICE
Deleting all test gas components deletes all
information of the test gas inlet.
NOTICE
It is necessary to specify the volume, volume type,
pressure and unit to use the gas manager.
Configured test gas
components
►
Enter volume
►
◄
Volume type
◄
Enter unit for
pressure
Enter pressure
►
Status
Test gas pressure
►
◄
State of test
gas pressure/
Gas cylinder level
indicator
NOTICE
Dräger recommends not exceeding a hose length of
10 m for the gas feed hoses.
i
i
i
i
i
i
Display
Gas pressure Meaning
<0.4 bar
Test gas pressure too low.
0,4 - 0,6 bar
Test gas pressure correct.
>0.6 bar
Test gas pressure too high.
Display
Gas cylinder level
61 - 100 %
30 - 60 %
< 30 %
Gas cylinder level indicator deactivated
Select test gas for X-am 2000
►
Fresh air inlet setting
►
Expert settings
►
Select test gas for PAC 7000 OV
►
Gas cylinder level indicator
►