Ventilation hose, test hoses, E-Vent test outlet and
oxygen adapter
●
Wash thoroughly in warm soapy water and rinse thoroughly
under running water.
Then
●
place in a solution of one part wine vinegar and two parts
water for approx. 30 minutes.
●
Rinse thoroughly under running water. Shake hoses to drain
off all water and dry thoroughly..
●
Dry parts thoroughly.
Water in the test hoses or in the oxygen adapter may impair
the operation of the unit.
Water in the ventilation hose or in the E-Vent test outlet can
reduce the air flow through the test outlet and allow the
CO
2
content of the inspiratory air to rise.
Assemble unit
see pages 10 to 12.
Atemschlauch, Meßschläuche, Meßblende E-Vent und
Sauerstoff-Adapter
●
In warmem Seifenwasser gründlich reinigen, unter
fließendem Wasser gründlich spülen,
danach
●
Einlegen in eine Lösung aus 1 Teil Weinessig und 2 Teilen
Wasser, ca. 30 Minuten wirken lassen.
●
Unter fließendem Wasser gründlich spülen. Wasser aus
den Schläuchen schütteln und die Schläuche gut trocknen.
●
Teile vollständig trocknen!
Wasser in den Meßschläuchen und im Sauerstoff-Adapter
kann die Funktion des Gerätes beeinträchtigen.
Wasser im Atemschlauch oder in der Meßblende E-Vent
kann das Abströmen von Luft an der Meßblende reduzieren
und den CO
2
-Gehalt in der Inspirationsluft ansteigen
lassen.
Gerät zusammenbauen
siehe Seite 10 bis 12.