Commissioning the test device
Dräger Quaestor 5000/7000
19
11. Tighten the high-pressure
outlet/L-piece screw
connection (6) using
a hexagon socket (wrench
size 14).
12. Tighten the L-piece/
200 bar module screw
connection (7) using a
hexagon socket (wrench
size 14).
13. Tighten all screw
connections on the
200 bar option.
14. Tighten the 2 screws on
the bracket.
Fitting the adapter for mask/helmet combinations
1. Use the drill supplied to drill a hole at each of the points
marked on the test head.
2. Align the adapter and
fasten with the two screws.
3.3.3
Setting up the test device
1. Align the table mount
parallel with the table edge.
The distance between the
table edge and the front
holes on the table mount
must be 12 cm.
2. Fasten the table mount to
the table using the screws
supplied.
3. Raise the test device
slightly at the back.
4. Align the front feet on the test device with the slots in the
table mount and move the test device backwards in the
table mount. The back feet must engage with the table
mount.
5. Ensure that the vent
screw (1) on the high
pressure inlet is closed.
6. Connect the compressed air pipework or the compressed
air hose to the high pressure inlet.
7. Connect the test device to the PC using the USB cable.
8. Connect the test device to the power supply:
Align the red marking on the power pack plug and on the
test device with one another, and insert the power pack
plug into the test device.
Connect the power supply unit to the mains.
NOTICE
Dräger also recommends connecting pressure
reducer R 57 584 to the high pressure input.
This makes it possible to ensure that the air is only fed
into the test device with a pressure of 200 bar.
NOTICE
The "top" of the test head is approx. 5 mm thick,
the test head is hollow inside. The interior of the test
head cannot be damaged by the drilling.
NOTICE
Position the test device on a table or workbench.
The PC should be positioned next to the test device so
that it is within the operator's reach.
01321868.eps
6
7
i
i
i
i
02321868.eps
i
i
WARNING
Compressed air pipework may only be connected by
trained personnel.
Failure to follow this warning could lead to death or se-
rious injury.
CAUTION
For the test device, only breathing air according to
EN 12 021 from a stationary compressed air network
supply or from a compressed air cylinder may be used.
If the test device is inflated with oxygen, there is a risk
of fire.
The maximum permissible high pressure is 300 bar
(4500 psi). The maximum permissible medium
pressure is 20 bar (58 psi). At higher pressures,
both the test device and the breathing safety product
may be damaged.
NOTICE
To disconnect the plug, move the socket approx. 1 mm
away from the test device. This releases the lock,
and the plug can be removed from the test device.
01521868.eps
12 cm
01621868.eps
1
!
!
i
i