Commissioning the test device
Dräger Quaestor 5000/7000
17
3
Commissioning the test device
3.1
Instructions on using the test device
To ensure safe and accident-free work with the test device,
observe the following points when working with the test device:
The test device must only be connected to the power
supply using the power pack supplied.
Before commissioning, the test device must be attached to
the workstation using the table mount.
The test device must be protected against ingress of water
and moisture. Ensure that no moisture from splash water or
leaking water can ingress into the test device.
The housing may only be opened by trained Dräger personnel.
The test device must be sent to Dräger for inspection once
per year. Keep the packaging supplied for future shipping.
3.2
Preconditions for use
The use of the test software requires proficiency in MS Windows.
The PC on which the test software is to be installed must meet
the following minimum requirements:
Processor: 1.6 GHz or higher
RAM: 2 GB or more
Free disk space: >500 MB
Operating system: Windows 7 or Windows 8
1 free USB port
The monitor resolution should be set to 1024 x 768.
3.3
Prior to first use
Before the test device can be put into operation, the following
steps must be carried out:
1. Install the test software on the PC and create the data
2. If applicable, mount accessories onto the test device
3. Set up the test device (see chap. 3.3.3 on page 19).
3.3.1
Installing the test software on the PC and creating
the data
1. Start "Protector.exe" from the USB flash drive.
2. Follow the instructions in the setup program.
The required program components will be installed in the
selected folder and an entry created in the start menu.
3. Observe the information displayed during installation.
4. Re-start the PC and log in.
5. Import the required device data with the test settings from
the USB stick. The following data sets are available:
For the EU database in the EU directory:
-
mask_EU.zip (apparatus data for full face masks)
-
LDV_EU.zip (apparatus data for lung demand valves)
-
reducer_EU.zip (apparatus data for pressure
reducers)
-
SCBA_EU.zip (apparatus data for compressed air
breathing apparatus)
-
BG4_EU.zip (apparatus data for closed-circuit
breathing apparatus)
-
CPS_EU.zip (apparatus data for chemical
protection suits)
-
SCUBA_EU.zip (apparatus data for diving equipment)
-
Saver_EU.zip (apparatus data for Saver PP)
-
Colt-Micro-ASV_EU.zip (apparatus data for Colt/
Micro/ASV)
For the NAFTA database in the US directory:
-
mask_US.zip (apparatus data for full face masks)
-
LDV_US.zip (apparatus data for lung demand valves)
-
reducer_US.zip (apparatus data for pressure
reducers)
-
SCBA_US.zip (apparatus data for compressed air
breathing apparatus)
-
BG4_US.zip (apparatus data for closed-circuit
breathing apparatus)
-
CPS_US.zip (apparatus data for chemical
protection suits)
In the test software, create additional users as well as
respiratory protection products, customers and locations
as required (see "First steps" in the help pages).
3.3.2
Fitting accessories
Fitting brackets for gauge or Bodyguard (Dräger
Quaestor 7000 only)
Screw the holder on the
right and left underneath
the test device using the
2 bolts supplied.
NOTICE
For installation, the user must have administrator
rights for the PC.
i
i
NOTICE
On delivery, the test software includes the following
users:
The user "Chief" with the password "Createdata"
The user "admin" with the password "admin"
Information on the test software is also provided on the
help pages.
i
i
00621868.eps