24
Quick Reference Guide Primus
Infinity Empowered
SW 4.5n
Alarms
The new alarm limit is now active. The cursor returns to
the
symbol.
Adapting alarm limits
When ventilation settings have been made, Primus
Infinity Empowered
can automatically adapt the alarm
limits for minute volume
MV
and the airway pressure
P
AW
to the current parameters in
Volume Mode
,
Vol-
ume AF
,
Pressure Mode
,
and
Pressure Support
(op-
tional).
z
Press the soft key
auto-set limits
(A).
The alarm limits menu opens automatically.
The alarm limits for
MV
and
P
AW
are adapted and high-
lighted by dark green background (B).
To quit the alarm limits menu:
z
Push the rotary knob or the
key.
The new alarm limits for
MV
are calculated by Primus
Infinity Empowered
from the measured value for the
minute volume
MV
in
Volume Mode
,
Volume AF
,
Pres-
sure Mode
, and
Pressure Support
(optional).
Setting the alarm tone
The volume of the alarm tone can be set in the configu-
ration menu.
Alarms in the Standby mode
All technical alarms, e. g. failure of equipment compo-
nents and a number of special operating states, are also
indicated by Primus
Infinity Empowered
when in
Stand-
by
.
A corresponding message appears in the alarms field,
but without any acoustic warning.
339
Volume
MV
6.0
3.6 8.4
P
AW
25
10 30
A
B