12
Quick Reference Guide Primus
Infinity Empowered
SW 4.5n
Operation Concept
The settings field highlighted in yellow returns to the
preceding menu level.
4
Select and confirm the
arrow (F) to exit the
menu.
Color concept
Colors are used to highlight operating sequences and
indicate the status of the soft keys.
System LED indicators
To display the status of the gas and power supply of the
Primus
Infinity Empowered
, several LED indicators are
located at the bottom of the screen.
A
Central gas supply (CS)
B
Reserve gas cylinders
C
Power supply
The Primus
Infinity Empowered
can be supplied via
the mains power or via a battery. The LED of the ac-
tive power supply lights up green.
361
Light green
– can be operated, leads to anoth-
er menu or operating function
– not yet active, presettings
Yellow
– selected, can be changed or set,
not yet confirmed
Dark green
– active parameter, can be operat-
ed
– current selection (configuration
menu)
Gray type
– cannot be operated
Standby Conf.
logbook entries
triggered by
time interval
F
yes
warning alarms
yes
no
caution alarms
no
1
5
10
2
COM 1
MEDIBUS
COM 2
MEDIBUS
Select
MEDIBUS
baud rate(k)
1.2
9.6
baud rate(k)
1.2
9.6
interfaces
logbook
MEDIBUS.X
MEDIBUS V4
370
LED lights up
green
CS line is connected and the pressu-
re is within the specified range
LED off
The pressure is not within the speci-
fied range or no CS line is connected
LED flashes
green
Malfunction of the corresponding in-
put pressure reducer
LED lights up
green
Reserve gas cylinder is connected
and the pressure is within the speci-
fied range
LED flashes
red
Reserve gas cylinder is connected,
but the pressure is not within the spe-
cified range and no central gas supp-
ly is connected.
LED off
Reserve gas cylinder is connected,
but the pressure is not within the spe-
cified range, the central gas supply is
connected and within the specified
range.
Or
no reserve gas cylinder connected.
O
2
Air N
2
O
O
2
Air N
2
O
A
B
C