Quick Reference Guide Primus
Infinity Empowered
SW 4.5n
Alarms
Alarms
Alarm displays
Alarm messages are displayed in the alarm field (A) in
order of priority.
All displayed alarms are sorted according to the three
classes. Within these classes, the alarms are sorted
and displayed according to an internal priority system. A
priority of 31 indicates the highest, a priority of 1 the low-
est priority. The priority numbers are given in the table
"Alarm messages". The internal priorities are not dis-
played.
Up to three messages can be displayed simultaneously.
In some cases, the corresponding measured values are
highlighted on the screen by a flashing background in
addition to the alarm message.
If more than three alarms occur simultaneously, the
symbol
appears (B) to the right of the alarm field and
the soft key
show all alarms
(C) is activated on the
right-hand side of the screen.
When this soft key is pressed, the upper curve display
(D) is replaced by up to six additional alarm fields for
15 seconds.
If more than nine alarm messages are simultaneously
active, the lowest priority alarms will not be displayed
until the total number of active alarm messages falls be-
low nine.
The alarm tone sequence accompanying a displayed
alarm message with the highest priority will always be
sounded at least once completely. The alarm tone se-
quences of alarm messages with lower priorities will not
sound if a higher priority alarm is activated, i.e. the tone
sequence thereof will sound.
If an alarm message of the same class as an active
alarm message is generated, the alarm tone of the new
alarm only sounds if the priority is higher than the prior-
ity of the previously active alarm.
NOTE
The Primus
Infinity Empowered
Quick Reference Guide is not a replacement or substitute for the Instructions for
Use. Any use of the device requires full understanding and strict observation of the Instructions for Use.