Dräger Polytron 8200/8310

Settings menu

8

Settings menu

This menu contains all the functions that are required for the

individual configuration of the device.

8.1

Switching SIL lock on or off

With this function, the device can be protected against

unauthorized changes to the configuration with a password.

Changing the configuration (e.g., changing the measurement

range from 100 %LEL to 50 %LEL) will cause the device to

display all the safety-relevant parameters to the user once more

before it returns to the measurement mode. The correctness of

the parameters and settings must be confirmed. 
1. Select 

Settings > Instrument > Set SIL

.

2. Select the desired state 

On

 or 

Off

 and confirm.

8.2

Device settings

8.2.1

Alarm settings (only with installed relay module)

Alarms issued via relays and LEDs.

Switching the alarm on or off

1. Select 

Settings > Instrument > Alarm > Alarm on/off

.

2. Select the desired state 

On

 or 

Off

 and confirm.

Configuring Relay A1 or A2

This function is used to define whether the alarm relay is

energized in standard operation or in an alarm state. If the

configuration of the relay is set to "Normally energized",

the corresponding relay contact is closed in the normal case

and drops out in the event of an alarm. This configuration

results in an alarm being triggered in the event of a loss of

power to the transmitter (fail-safe). The fault relay is factory-

configured to "Normally energized" and cannot be changed.

1. Select 

Settings > Instrument > Alarm > Relay A1 

or

Relay A2 

and confirm.

2. Select 

Normal energy supply

 oder 

Energy supply

during alarm

 and confirm.

Configuring Alarm A1 or A2

1. Select 

Settings > Instrument > Alarm > Alarm A1 

or

Alarm A2 

and confirm.

2. Select the line for entering the alarm limit and confirm.
3. Set the threshold and confirm.

The setting for the alarm threshold is displayed. 

4. Select 

Next

 and confirm.

The setting for the alarm direction is displayed.

5. Select 

Rising

 or 

Falling

 and confirm.

6. Select 

Next

 and confirm.

The setting for the alarm direction is displayed.

7. Select 

Latching

 or 

Non Latching

 and confirm.

8. Select 

Next

 and confirm.

The setting for the acknowledgement is displayed.

9. Select 

Acknowledgeable

 or 

NonAcknowledgeabl

 or

PreAcknowledgeabl

 and confirm.

10. Select 

Next

 and confirm.

The function for setting the A1 hysteresis is opened.

11. Select the line for entering the hysteresis and confirm.
12. Set the hysteresis for A1 and confirm.

The hysteresis setting for A1 is displayed. 

13. Select 

Next

 and confirm.

The settings for A1 are displayed.

14. Select and confirm with 

Confirm

.

The settings for the A1 alarm are now complete.

NOTICE

SIL lock is only available for certain sensors (see also

the sensor data sheets or the Dräger Polytron 8X00

Safety Manual). 

NOTICE

When SIL is activated, the device displays all the

important parameters and setting before returning to

measurement mode. Check the parameters and

settings carefully and confirm.

CAUTION

In the normal state, the relays in the relay module are 

energized. That is, the relays change their state if the 

power supply is interrupted!

CAUTION

If the alarm is inactive, the alarm state will not be

issued by the LEDs or the relay interface! A warning

message   will be shown on the display!

i

i

i

i

!

!

NOTICE

This function permits the setting of a zone within which

a triggered relay maintains its status until the gas

concentration is outside the zone. This function prevents

relays from chattering at an alarm threshold. 

Example: A2 at 40 %LEL methane

Maximum possible hysteresis: 3 %LEL

Alarm activation at measured values 

40 %LEL

Alarm deactivation at measured values 

37 %LEL

(40 %LEL - 3 %LEL)

i

i

Содержание Polytron 8200

Страница 1: ...structions for Use The user must fully understand and strictly observe the instructions Use the product only for the purposes specified in the Intended use section of this document Dr ger Polytron 820...

Страница 2: ......

Страница 3: ...to the Quick Menu mode 13 4 5 Switching to Menu mode password protected 13 4 6 Measurement mode 13 5 Calibration 14 5 1 Calibrating the device 14 5 2 Performing Auto Calibration 15 6 Menu overview 16...

Страница 4: ...cording to Annex II Sections 1 5 5 1 5 6 and 1 5 7 of Directive 94 9 EG is not covered at present z Exchanging components may compromise the intrinsic safety This only applies if the device is intrins...

Страница 5: ...tronics and optional relay 4 Lower part of housing 5 Sensor 6 Assembled device 2 1 2 Explosion proof device with e Box for increased safety 1 Housing cover 2 Clip 3 Enclosure with main electronics and...

Страница 6: ...the concentration is performed by internal electronics and software For the output signal the gas sensor emulates the half bridge of a catalytic bead sensor 2 2 3 Contaminating gases for the Dr gerSen...

Страница 7: ...ipulations that may affect the installation of the system z For installation without a conduit an approved cable gland e g Hawke A501 421 A NPT or equivalent must be used see chapter 12 on page 28 To...

Страница 8: ...ing to the application e g HART operation certain minimum impedances must be adhered to see chapter 3 3 on page 8 The conductors for the power supply must have an adequately low resistance to insure t...

Страница 9: ...connected to the 9 way plug 1 Disconnect the 9 way plug on the rear of the main electronics 2 Connect the wires for Alarm Relay 1 Alarm Relay 2 and the Fault relay to the corresponding terminals see...

Страница 10: ...g SW dongles are available for the Polytron 8200 To install a SW dongle 1 Switch off the power to the unit or declassify the area according to the local regulations 2 Release the set screw and unscrew...

Страница 11: ...e chapter 4 2 2 on page 12 z SIL when SIL status is activated When the optional relay PCB is used z If the first alarm has been triggered the red LED gives single flashes z If the second alarm has bee...

Страница 12: ...imum measurement limit and measurement units for the sensor z Measurement range and measurement units for the analog interface Only displayed when the 4 20 mA interface is enabled Warning message pres...

Страница 13: ...g the video LED test software version and initialization and starts the warm up phase The display indicates that the sensor will be ready to operate in hh mm ss the countdown begins and the Polytron 8...

Страница 14: ...ars 2 Apply zero gas synthetic air or nitrogen to the sensor 3 Select Next and confirm The current value is displayed After the displayed value is stable 4 Set the value to 0 and confirm The message P...

Страница 15: ...ration is started The calibration gas concentration the units and where appropriate the calibration gas are displayed and can be changed 4 Start the calibration with Next Back to menu aborts the calib...

Страница 16: ...18 Device init page 20 Set alarm A1 page 19 Datalogr on off SW dongle page 20 Set alarm A2 page 19 Set Datalogger Display page 20 Set fault page 19 Clear datalogr Calibration PWD page 19 Sample time S...

Страница 17: ...f all the numerical groups are displayed as 00 there are no faults or warnings present 7 1 4 Displaying installed modules Display of the installed hardware modules z Select Information Instrument Modu...

Страница 18: ...ering the alarm limit and confirm 3 Set the threshold and confirm The setting for the alarm threshold is displayed 4 Select Next and confirm The setting for the alarm direction is displayed 5 Select R...

Страница 19: ...the function key This function is used to assign a predefined function to the key The selected function is activated by briefly taping the key 1 Select Settings Instrument Function key and confirm 2 S...

Страница 20: ...ion 1 Select Settings Instrument Display Displaytest and confirm 2 Select On or Off and confirm When this function is activated the display is shown inverted and all the LEDs are lit 8 3 Communication...

Страница 21: ...line and confirm 3 Set the currrent and confirm 4 Select Confirm and confirm Switching the warning signal on or off This function is used to switch the warning signal on the analog interface on and o...

Страница 22: ...selected concentration 1 Select Settings Communication Analog interface Set concentr and confirm 2 If the alarm in the control unit is disabled confirm the message Disable all alarm settings 3 Select...

Страница 23: ...w of the new gas settings is displayed z Return to the previous view with Previous or confirm the settings with Confirm Information about measurement units The Polytron 8200 8310 represents the measur...

Страница 24: ...percentage of the full scale reading Example A trigger threshold of 2 is entered for a full scale reading of 500 ppm Consequently the datalogger will only store measurements if they deviate relative t...

Страница 25: ...device checked by Dr gerService 10 4 20 mA interface cable not connected Disconnect the powwer supply connect the 4 20 mA interface and reconnect the power supply 045 Device not detecting any sensor...

Страница 26: ...occurs again Have the device checked by Dr gerService 105 Device fault Have the device checked by Dr gerService 104 No valid date or time set Set time and date see chapter 8 2 3 on page 19 107 Batter...

Страница 27: ...he set screw and unscrew the housing cover from the device 3 Lift the handle and remove the enclosure with the main electronics 4 Remove the protective caps for the connectors 5 Unscrew the connector...

Страница 28: ...or Ex DD 105 mA at 24 V DC no relay sensor not remote 145 mA at 24 V DC with relay remote sensor with Dr gerSensor Ex LC 130 mA at 24 V DC no relay sensor not remote 165 mA at 24 V DC with relay remot...

Страница 29: ...ue Nm Housing cover min 266 min 30 Sensors min 266 min 30 Plug min 266 min 30 Cable bushing min 443 min 50 Device to e Box 71 8 Electronic Torque Lb In Cable size in AWG Cable size in mm2 All field wi...

Страница 30: ...A2 alarm is present If an alarm is configured as acknowledgeable and the alarm is acknowledged the LED lights Alarm hierarchy A2 has a higher priority for the LED than A1 However the A1 and A2 relays...

Страница 31: ...ote LC e S 4 20 HART 45 44 538 Dr ger Polytron 8200 Remote LC e S 4 20 HART Relay 45 44 539 Dr ger Polytron 8310 d A 4 20 HART 45 44 807 Dr ger Polytron 8310 d A 4 20 HART Relay 45 44 808 Dr ger Polyt...

Страница 32: ...PE Europe 68 06 978 Calibration adapter Viton 68 10 536 Process adapter stainless steel with M30 x 1 5 locking nut for Dr gerSensor DD 68 12 470 Process adapter stainless steel with M36 x 1 5 locking...

Страница 33: ...Declaration of Conformity Dr ger Polytron 8200 8310 33 14 Declaration of Conformity...

Страница 34: ...www norrscope com...

Страница 35: ......

Отзывы: