Installation Procedure for Resuscitaire® Radiant Warmer Model RW82
1.
Place the radiant warmer flow label (A) (MU11144, RW82) on the unit directly above
the existing flow label (see figure 1 on page 6).
2.
Install the micro-volume suction bottle holder (C) on the existing suction bottle holder
clips (B).
NOTE:
If it does not fit securely, pinch the clips together and release.
3.
Mount the snap-in wire tie (I) on the door using the shoulder washer (J) and hex nut
(K) provided.
4.
Pull the supply connection tubing (D) toward the front of the unit, and mount the tub-
ing in the supply connection clamp (L).
5.
Connect the micro-volume suction bottle filter (F) in line with the supply connection
tubing (D) and the suction bottle (G).
6.
Place the suction bottle (G) in the micro-volume suction bottle holder (C).
7.
Secure the supply connection tubing (D) with the snap-in wire tie (I).
8.
Connect the desired suction tubing to the outlet port (H) of the suction bottle cap, and
secure the free end of the extension tubing in either tubing retaining slot provided on
the front or rear panel of the bassinet.
Figure 1. Resuscitaire® Radiant Warmer Model RW82