Technical manual
|
X-am
®
2800
35
Device settings
8.2
Device and sensor settings
LEL factor
2)
CH
4
(methane)
3)
H
2
(hydrogen)
C
3
H
8
(propane)
4.4 Vol% (corresponds to 100 %LEL)
4.0 Vol% (corresponds to 100 %LEL)
1.7 Vol% (corresponds to 100 %LEL)
STEL
STEL feature – inactive; average dura
-
tion = 15 minutes
TWA
TWA feature – inactive; average dura
-
tion = 8 hours
Configuration type alarm thresholds
ATEX compliant
Alarm A1
Acknowledgeable; non-latching, pre-
alarm, increasing measured value (for
O
2
sensor, decreasing measured value
as well)
Alarm A2
Not acknowledgeable; latching, main
alarm, increasing measured value (for
O
2
sensor, decreasing measured value
as well)
Expired bump test interval
Channel warning
Expired calibration interval
Channel warning
1) The set capture range can be read out on the gas detector and activated or deactivated. The
capture range is activated in measuring mode ex works. The capture range is always deacti
-
vated in calibration mode.
2) An LEL factor can be adapted to national regulations using the Dräger CC-Vision PC software.
3) Notation in the gas detector: ch4
Feature
Setting
Name:
Area / setting
Device settings:
Password(s)
Numerical range (4-digit)
Acoustic operation signal
Yes / No
Switch-off mode
“Switch off is allowed” or
“Switch off not allowed” or
“Switching off not allowed during A2”
Short-term exposure limit (STEL)
1)2)
0 - 60 (in minutes; setting for exposure
alarm)
Shift length (TWA)
3)
60 - 1440 (in minutes; setting for expo
-
sure alarm)
Sensor settings:
A1 alarm:
Latching
Acknowledgeable
On / Off
On / Off
A2 alarm:
Acknowledgeable
On / Off
Alarm threshold A1 increasing
(in measurement module)
0 to A2