Technical manual
|
X-am
®
2800
17
Troubleshooting
5.2
Warnings
The sensor technology that can be used depends on the device type.
The first digit of the code represents the sensor channel:
x01
No valid zero-point calibration.
Carry out a fresh air calibration /
zero-point calibration.
x02
No valid span calibration.
Carry out the span calibration.
x03
Measured value of sensor in neg
-
ative range.
Carry out a fresh air calibration /
zero-point calibration.
x04
Sensor not connected or faulty.
Connect sensor or check contacts.
x05
Error during bump test.
Repeat the bump test.
x07
Faulty rise time test.
Repeat the rise time test with X-
dock.
x08
Invalid value
x09
Channel error
Contact DrägerService.
x12
Calibration interval has expired.
Carry out the span calibration.
x13
Measurement invalid.
Reinaugurate sensor.
x14
Sensor hardware error
Replace the sensor.
x19
If a channel error occurs on the
CO-channel when activating Tox
-
icTwins, this error will be set on
the HCN channel.
Recalibrate the CO-channel or
replace the sensor.
x21
Insufficient oxygen for the correct
functionality of the CatEx sensor.
Operate the gas detector in atmo
-
spheric conditions.
x22
Zero calibration failed (CatEx ther
-
mal conduction range).
Carry out a fresh air calibration /
zero-point calibration.
x23
Span calibration failed (CatEx
thermal conduction range).
Carry out the span calibration.
x24
Gas detector incorrectly config
-
ured by Dräger CC-Vision.
Replace sensor for affected chan
-
nel with Dräger CC-Vision.
x36
Device temperature too high.
Operate gas detector within per
-
mitted temperature range.
x37
Device temperature too low.
Operate gas detector within per
-
mitted temperature range.
216
Thermal conduction calibration
interval for DrägerSensor CatEx
SR has expired.
Perform span calibration for
DrägerSensor CatEx SR.
218
Blocking alarm not plausible.
Calibrate sensor.
242
Sensor current for DrägerSensor
CatEx SR too low.
Contact DrägerService.
Code
Cause
Remedy
Code
Sensor channel
Code
Sensor channel
2xx
DrägerSensor CatEx SR
5xx
EC3