30
Technical manual
|
X-am
®
2800
Maintenance
Procedure:
1. Connect the gas detector to a PC. For further information, see the following
chapter: "Configuring the gas detector with the PC and reading the data
2. Start the sensor exchange wizard in the Dräger CC-Vision PC software and
follow the instructions.
3. Deactivate the corresponding sensor slot.
4. Open the gas detector.
a. Loosen the screw and remove the power pack.
NOTICE
Damage to the display
When lifting out the front housing cradle, it can happen (in rare cases) that the
display remains caught in the front housing cradle and becomes damaged as a
result.
► Lift out the front housing cradle carefully. If the display gets caught, first
carefully loosen the display from the front cradle and then remove the front
housing cradle entirely.
b. Loosen 4 screws and carefully remove the front housing cradle.
5. Exchanging the EC sensors:
51586
51588