Instructions for use Babyleo TN500 SW 1.0n
23
Safety-related information
2.16 Safety
instructions for accessories
2.16.1 Accessories
The approved accessories for this product are included in the list of accessories
and listed in declarations by Dräger (e.g., declaration of conformity). The use of
other accessories may adversely affect the functional integrity of the product and
lead to personal injury and property damage.
►
Only use approved accessories.
►
Only installation personnel are authorized to mount accessories.
►
Observe the current list of accessories and the assembly instructions of the
device.
►
Observe the instructions for use and assembly instructions for the accessories.
►
Observe the weight limit for attached accessories.
►
When attaching accessories, make sure that the device does not tip over.
2.16.2 Ventilation
The temperature measurement for breathing gas humidification may be distorted by
the radiant warmer or the incubator temperature. There is a risk of hypothermia if
the breathing gas is too cold and of suffocation due to condensate formation.
►
Follow the instructions for use of the humidifier in conjunction with the
instructions for use of the breathing hose.
2.16.3
Hoses and cables
If the device or the trolley is moved in an uncontrolled manner, patient treatment
may be interrupted. The patient is put at risk.
►
When using the height adjustment mechanism and the trolley, always make sure
that connected hoses and cables are not accidentally disconnected.
►
Lock the double castors on the trolley with the locking brakes.
►
Only use hoses and cables of sufficient length.
►
Route hoses and cables into the patient compartment through the hose
grommets.
►
Do not run hoses and cables over the hood arm.
2.16.4
Objects in the patient compartment
If objects are placed in the patient compartment, the patient may be infected by
pathogens. The patient may swallow parts and choke.
►
Do not place any objects in the patient compartment.
2.16.5 Working
light
Conducting a medical examination of the patient without the use of lighting with
neutral colors may cause e.g., misinterpretation of the skin color.
►
Do not use the working light on the device when examining the patient.
►
Always use an examination light in accordance with the standard IEC 60601-2-
41 when examining the patient.
Содержание Babyleo TN500
Страница 2: ...2 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 212: ...212 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 214: ...214 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 221: ...Instructions for use Babyleo TN500 SW 1 0n 221 This page has been left blank intentionally...