144
Instructions for use Babyleo TN500 SW 1.0n
Configuration
8.4.1.4 Horizontal
tab
Alarms
Vertical tab
Heading or parame-
ter
Description
Factory setting
Air mode/Man.
mode
Enter and confirm the configu-
ration password.
Upper limit for air
temp. dev
Specify the upper alarm limit
for the air temperature devia-
tion.
+ 1.5 °C (2.7 °F)
Lower limit for air
temp. deviation -
Specify the lower alarm limit for
the air temperature deviation.
- 1.5 °C (2.7 °F)
Advanced skin tem-
perature monitoring
In air temperature mode and
manual mode, the patient's
skin temperature can be moni-
tored. The skin temperature
sensors must be connected to
the device and attached to the
patient.
Off
If the
Advanced skin tem-
perature monitoring
function
is activated, the following
alarm limits can be defined:
–
The upper alarm limits for
the central and the periph-
eral skin temperature
Skin temperature
39 °C (102.2 °F)
–
The lower alarm limits for
the central and the periph-
eral skin temperature
Off
–
The upper alarm limit for
the difference between the
central and peripheral skin
temperature
5 °C (9 °F)
–
The lower alarm limit for the
difference between the cen-
tral and peripheral skin tem-
perature
Off
Содержание Babyleo TN500
Страница 2: ...2 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 212: ...212 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 214: ...214 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 221: ...Instructions for use Babyleo TN500 SW 1 0n 221 This page has been left blank intentionally...