Instructions for use Babyleo TN500 SW 1.0n
181
Service
12.2.2 Verification
of
the NAWI scale (option)
Details regarding the conformity assessment carried out by the manufacturer and
the calibration made prior to installation at the operating location can be found on a
product label. In the stated area the scale can be operated with the specified
measurement uncertainty. For further information, see the following chapter:
''Technical data'' (page 188)
The verification is no longer valid once the verification deadline has passed. The
verification is also invalid in the following situations:
–
After modifications of the basic scale settings
–
Illegible or damaged main stamp or security stamp
–
After changes in the service dialog due to pressing the
Adjust
(Adjust) button
–
Exceeding the service error limits
In these situations, a new official verification in accordance with national regulations
is required. Verification of the scale may only be carried out by a local measuring
body and the relevant verification authority. To check that the scale is accurate and
is working correctly, specialized service personnel can carry out an appropriate
examination of the scale. For more information, contact DrägerService.
If the scale was installed by Dräger, the conformity assessment has already been
carried out. Before the verification deadline expires, the scale must be verified in
accordance with national regulations.
12.2.3 Safety
checks
Safety checks are no substitute for maintenance including preventive replacement
of wearing parts as specified by the manufacturer.
12.2.3.1
Performing the safety checks
WARNING
Risk due to missing service measures
If the device is not maintained at the prescribed intervals, the function and the
safety of the device cannot be ensured.This may put the health of patients and
users at risk.
►
Perform service measures on the device at the prescribed intervals.
►
Service measures must only be performed by suitably qualified persons.
WARNING
Risk of patient injury due to service measures performed during therapy
Performing service measures during therapy may endanger the patient.
►
Only perform service measures when there is no patient in the device.
Содержание Babyleo TN500
Страница 2: ...2 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 212: ...212 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 214: ...214 Instructions for use Babyleo TN500 SW 1 0n This page has been left blank intentionally...
Страница 221: ...Instructions for use Babyleo TN500 SW 1 0n 221 This page has been left blank intentionally...