25
WARNING
Fault in display
Cause
Remedy
"Accuracy check overdue lockout in X
days"
Accuracy check intervall expired. Device is
useable until graceperiode is expired.
Download and clear datalogger. Perform
accuracy check.
"Service required
Accuracy check"
Accuracy period expired.
Download and clear datalogger. Perform
accuracy check.
"Service date expired / accuracy check"
Service date expired
Download and clear datalogger. Perform
accuracy check.
"Calibration overdue lockout in X days"
Calibration intervall expired. Device is usea-
ble until graceperiode is expired.
Perform calibration by trained service person-
nel, e. g. by Dräger Safety.
"Service required
Calibration"
Calibration period expired.
Perform calibration by trained service person-
nel, e. g. by Dräger Safety.
"Service date expired / calibration"
Service date expired
Perform calibration by trained service person-
nel, e. g. by Dräger Safety.
"Memory for XXTests"
Device only useable for X tests.
Transfer memory data to PC. Clear memory.