Dräger Alcotest 5510
35
Technical data
Intervals before readiness for operation
1st measurement
about 6 s
after terminating a result of a previous test in range:
(breath alcohol unit)
0.25 mg/L
1
20 s
0.50 mg/L
1
40 s
1.00 mg/L
1
80 s
> 1.5 mg/L
120 s
after terminating a result of a previous test in range:
(blood alcohol unit)
0.50 ‰
1
20 s
1.00 ‰
1
40 s
2.00
‰
1
80 s
> 3.00 ‰
120 s
Intervals for other measuring units selected can be calculated
accordingly (see page 34).
Calibration
Calibration interval
depends on required accuracy,
typically: 6 months
Permissible calibration range for unit of measuring value:
mg/L
0.160 to 0.750 mg/L
µg/L
160 to 750 µg/L
g/210L
0.035 to 0.150 g/210L
µg/100 mL
16 to 75 µg/100 mL
µg%
16 to 75 µg%
‰
0.35 to 1.50 ‰
%
0.035 to 0.150 %
g/L
0.35 to 1.50
g/L
mg/100 mL
35 to 150
mg/100 mL
mg%
35 to 150
mg%
mg/mL
0.35 to 1.50
mg/mL
Standard calibration value when using Dräger Ethanol-Standard
solution for unit of measuring value:
mg/L
0.476 mg/L
µg/L
476 µg/L
g/210L
0.100 g/210L
µg/100 mL
47.6 µg/100 mL
µg%
47.6 µg%
‰
1.00
‰
%
0.100
%
g/L
100
g/L
mg/100 mL
100
mg/100 mL
mg%
100
mg%
The use of these standard calibration values corresponds to a
conversion factor of 2100 between breath and blood alcohol
concentration-values.
Dimensions
142 mm x 75 mm x 36 mm (H x W x D)
Weight
approx. 195 g
Electrical supply
Batteries
2 x 1.5 V Type Alkaline
(Mignon, LR6, AA)
approx. 1500 measurements
CE marks
Electromagnetic compatibility
(Directive 89/336/EEC)