41
Pump operation is not functioning.
Incorrect holder (diffusion) in
position.
Position the holder (pump).
Pump fault during operation.
Condensate formation with cold and
damp intake air.
Use hose set with water trap
(order no. 83 21 527).
Pump outside the specified range.
Have the pump checked by Dräger.
Flow test has failed.
Flow test has not been performed.
Perform flow test, repeat if
necessary.
Holder (pump) is not correctly
positioned.
Re-position holder (pump) and
check that it is correctly fitted.
Horn is too quiet.
Volume is set too low.
Set the volume using the Dräger
CC-Vision PC software (refer to
“Configuring the device” on
page 39).
Optical alarm signals not visible or
poorly visible.
Configuration or pattern incorrectly
set.
Configure the optical alarm signals
using the Dräger CC-Vision PC
software (refer to “Configuring the
device” on page 39).
Life signal is not functioning.
Configuration is set incorrectly.
Configure the life signal using
the Dräger CC-Vision PC software
(refer to “Configuring the device” on
page 39).
Gas detection instrument not
detected.
IR interface dirty.
Clean IR interface.
Incompatible gas detection
instrument.
Use Dräger X-am 5000.
Incorrect software version in the gas
detection instrument.
Have a software update performed
by Dräger.
IR interface faulty, clip on gas
detection instrument not correctly
positioned.
Close clip on Dräger X-am 5000.
Holder is not correctly positioned.
Re-position holder and check that
it is correctly fitted.
Dräger X-am 5000 switches off
quickly; no power supply.
Power contacts dirty or damp.
Clean power contacts.
Device defect displayed.
Dräger X-am 5000 moved away
from Dräger X-zone 5000 during
operation.
Acknowledge alarm on
Dräger X-zone 5000, switch off
Dräger X-zone 5000.
Fault
Cause
Remedy