Phototherapy
To reduce risk of hyperthermie you may use photo-
therapy in combination with the RH 600 infant radiant
heater.
Otherwise, risk of inducing hypothermia in the undressed
infant.
Example: Wall-mounted RH 600 with Babytherm 8000 WB
and Photo-Therapy 4000
●
Push the stand of the phototherapy unit under the
frame of the infant bed, e.g. Babytherm 8000 WB.
●
Tilt the lamp housing towards the bed.
●
Adjust the height of the phototherapy unit: minimum
clearance between the lower edge of the lamp body
and the patient: 0.3 metres.
●
Position the lamp housing so that the radiant heat
reaches the patient unimpeded.
●
Follow the specific Instructions for Use of the
phototherapy units.
If operating the radiant heater in »Man« mode:
●
Reduce the heater output level by 3 steps
compared to operation without phototherapy.
Using the phototherapy units will result in increasing the
temperature of the infant.
Monitor the core temperature of the infant.
47
Phototherapy
min.
0,3 m
Fuse
5
6
7
Содержание 2M 20 116
Страница 5: ...Montagem glichkeiten Montage an einem Fahrgestell Wandmontage 5 Zweckbestimmung 1 382 94...
Страница 27: ...Abmessungen Strahler Strahler auf Fahrgestell 27 Technische Daten 500 1200 bis 1770 920 100 205 200 685 60...
Страница 33: ...Installation options Mounted on wheeled stand Wall mounted 33 Intended Use 1 382 94...
Страница 61: ...Possibilit s de montage montage sur un chariot montage mural 61 Domaine d application 1 382 94...
Страница 89: ...Posibilidades de montaje Montaje sobre soporte m vil Montaje mural 89 Aplicaci n 1 382 94...
Страница 114: ...114...
Страница 115: ...115...