![Dräger 2M 20 116 Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/drager/2m-20-116/2m-20-116_instructions-for-use-manual_2526506041.webp)
1 Press the power switch until it clicks into position =
ON.
After the self-test, the apparatus switches to »Man«
(manual) operating mode.
If the temperature is outside the measuring range from
33
°
C to 38
°
C:
●
3 dashes at bottom of display =
Temperature below 33
°
C.
●
3 dashes at top of display =
Temperature above 38
°
C.
●
See section on "Troubleshooting", page 50.
Allow at least 5 minutes for the skin temperature
sensor to adjust to the temperature of the infant.
If the displayed value remains constant:
2 Press the
¡
key: skin temperature control is
activated, and the green LED is lit.
3 The skin temperature is displayed in the temperature
display.