For your safety and that of your patient
9
Product-specific safety information
CAUTION
The general state of the device and internal cables and
hoses has to be checked for damages.
For proper function of the device it must be examined
carefully before use.
Functional safety
This device does not have any Essential Performance
features as defined by IEC 60601-1-2, because there is no
active electronic equipment in the device.
For that reason there are no effects concerning electrical
irradiation and radiation.
WARNING
Components that have been subsequently installed on
the medical device must correspond to all regulatory or
technical requirements and approvals.
The customer must make the right choice. Dräger
assumes no warranty and no liability for parts from third-
party manufacturers or the functionality of third-party
manufacturers' products.
Only trained personnel may carry out subsequent
installation.
If the above are not complied with, the correct functioning
of the medical device may be compromised.
WARNING
Danger of electric shock! The device housing must
not be opened.
WARNING
Increased fire risk!
Gas terminal units must not come into contact with
oil, grease, or flammable liquids.
WARNING!
The electrical connections from the hospital wiring
to the terminal strip must be made by a qualified
electrician.
WARNING!
Do not use this device in oxygenated areas.
NOTE
Certifications depend on market-specific components
(e. g. gas terminal units and power sockets).
CAUTION
Danger of personal injury and/or equipment damage
The supply unit must be in flawless operation condition.
Damage of the system can cause objects (defective gas
terminal units and power sockets, loose paint particles,
adhesive labels, etc.) to fall down into the operation field
and compromise the supply to the patient or the connected
devices.
CAUTION
Risk of equipment damage or personal injury
If a light is attached directly under a support arm of a supply
unit, the light will move when the support arm is positioned.
It may collide with people or objects as a result of this.
Avoid collisions when positioning the support arm.