
7
Accessories
WARNING
Only the accessories indicated in the chapter
"List of Accessories" on page no 31 have been
tested and approved for use with the medical
device.
Therefore, it is strongly recommended that
only these accessories are used in conjunc-
tion with the medical device. Otherwise, the
correct functioning of the medical device may
be compromised.
WARNING
If you push out the upper smaller pole of the
infusion pole too far, it may fall off under load.
Connected devices
WARNING
Risk of electric shock and of device malfunc-
tion
Any connected devices or device combina-
tions not complying with the requirements
mentioned in these instructions for use may
compromise the correct functioning of the
medical device. Before operating any combi-
nation of devices, refer to and strictly comply
with the instructions for use for all connected
devices and device combinations.
Safe connection with other electrical
equipment
Connection to other devices
Device combinations approved by Dräger (see
instructions for use of the individual devices) meet the
requirements of the following standards:
– IEC 60601-1 (EN 60601-1, 3rd Edition)
Medical electrical equipment. Part 1:
General requirements for safety
– IEC 60601-1-2 (EN 60601-1-2)
Medical electrical equipment
Part 1-2: General requirements for safety
Collateral standard: Electromagnetic compatibility;
Requirements and tests
– IEC 60601-1-6 (EN 60601-1-6)
Medical electrical equipment
– Part 1-6: General
requirements for safety
–Collateral standard:
Usability
If Dräger devices are connected to other Dräger
devices or third-party devices and the resulting
combination is not approved by Dräger, the correct
functioning of the devices may be compromised.
The owner is responsible for ensuring that the
resulting system meets the requirements of the
applicable standards.
Strictly observe assembly instructions and instruc-
tions for use for each networked device.
CAUTION
Risk of patient injury
Electrical connections to equipment not listed in
these instructions for use or these assembly
instructions must only be made when approved
by each respective manufacturer.