20
Please refer to the table below to troubleshoot any problems you may encounter when using the
nebulizer.
Problems
Possible Causes
Solutions
Extremely low nebulization.
Medication chamber is not
completely attached.
Re-attach the medication chamber
correctly and restart the power. (See
page 6)
No contact between
medication and mesh for
more than 15 seconds.
Adjust the nebulizer’s angle so the
medication can come in contact.
Mesh of medication
chamber is clogged.
Clean the medication chamber. If it
still cannot be used after cleaning,
please replace with a new medication
chamber.
Electrodes on medication
chamber are clogged with
medication or water.
Clear the electrodes of clogged
medication or water and restart the
power. (See page 14)
Electrodes on nebulizer and
medication chamber are
stained.
Remove the stains and restart the
power. (See page 14)
After turning power on, power
indicator lights for one second
and then immediately goes out.
Medication chamber is not
completely attached.
Re-attach the medication chamber
correctly and restart the power. (See
page 6)
No medication in
medication chamber.
Put in the medication in the
medication chamber. (See page 9~10)
No contact between
medication and mesh.
Adjust the nebulizer’s angle so the
medication can come in contact. (See
page 11)
Electrodes on nebulizer and
medication chamber are
stained.
Remove the stains and restart the
power. (See page 14)
Power indicator is not lit and
nebulizer is not nebulizing.
Batteries installed
backwards.
Re-install the batteries in the correct
orientation and restart the power.
(See page 7~8)
Low battery power.
Replace with new batteries and restart the
power. (See page 7~8)
Incorrect connection of AC adapter
to nebulizer.
Re-connect in the correct manner and
restart the power. (See page 8)
Power indicator is
lit and nebulizer is
not nebulizing.
Low-power indicator is lit
constantly, insufficient battery
power, or battery has run out.
Replace with new batteries and restart the
power. (See page 7~8)
Rupture of mesh of medication
chamber.
Replace with a new medication chamber and
then put in the medication. (See page 16)
Electrodes on medication chamber
Clear the electrodes of clogged medication
Troubleshooting