D.2.1.5 “Device parameters” panel 

D.2.1.5.1 The  “Device parameters” panel contains  information about static parameters, which 

are  not  used  in  the  measurements  process,  and  dynamic  parameters,  which  define  measurement 
characteristics of the device, and also general information about the device operation. 

D.2.1.5.2 Static parameters: 
-

 

Device address; 

-

 

Device; 

-

 

Device code; 

-

 

Firmware version; 

-

 

Configuration date; 

-

 

Assembly date; 

-

 

Device date/time; 

-

 

Volume level; 

-

 

Current device language; 

-

 

Factory number: 

Parameter  “Configuration  date”  is  a  variable;  it  is  automatically  changed  when  the  “Set 

parameters” button is pressed. 

Parameter “device address” represents the net address of the device. 
D.2.1.5.3 Dynamic parameters 
Each device has its own set of dynamic parameters. Some parameters can possess the value of 

0 or 1; some parameters can possess the value in the exponential form, for example 2.3E-003 (0.0023). 
Some parameters are “read only”, for example “Running hours”. 

Correction of the parameters is carried out as follows: 
-

 

double click on the line of the parameter to be corrected; 

-

 

in the column “New value” input new parameter value; 

-

 

press “ENTER” or click on any other line; 

-

 

if necessary, correct other parameters; 

-

 

press the button “Set parameters F2”; 

-

 

check  correctness  of  the  writing  –  new  parameter  values  will  be  displayed  in  the  column 

“Parameter point” after a while. 

ATTENTION!

  DECIMAL  POINT  (SEPARATOR  BETWEEN  INTEGER  PART  OF  THE 

NUMBER  AND  THE  FRACTION)  IS  USALLY  A  DOT  (.).  HOWEVER  YOUR  OPERATION 
SYSTEM  CAN  BE  SET  TO  USE  COMMA  (,)  INSTEAD.  BE  CAREFUL  WHEN  YOU  INPUT 
NUMBERS. 

Sets of values or ranges of values of parameters are indicated in the operational documentation 

for each device. 

D.2.1.5.4 Buttons on the “Device parameters” panel 
In the upper part of the “Device parameters” panel the following buttons are located: 
-

 

“DEVICE DATE/TIME” – this button is activated for devices that has or emulates built-in 

clocks; 

-

 

“SET DEVICE LANGUAGE OF THE INTERFACE” – this button is activated for devices 

with multilanguage interface, it allows setting the active language of the device; 

-

 

“SET  VOLUME  LEVEL”  –  this  button  is  activated  for  devices  equipped  with  adjustable 

signaling features, it allows to set the desired volume and, if necessary, to test it; 

-

 

“SET  ADDRESS”  –  in  the  broadcast  mode  this  button  allows  to  set  the  net  address,  for 

setting the address press the button “Set address”, then in the window that will appear input the new 
address  –  three  dot  separated  groups of  digits  (3  digits  in  each  group).  The  range  of  values  in  each 
group – from 002 to 254. After completion the input click “Ok”. The new address will be written into 
the non-volatile memory of the device. At the same time parameter “date of correction of parameters” 
is automatically set to current date; 

Содержание BDMN-200

Страница 1: ...Scientific Production Company Doza DETECTORS BDMN 200 User Manual AJAH 418266 007RE ...

Страница 2: ... of a product 8 2 4 Adjustment 8 3 Maintenance 8 3 1 General notes 8 3 2 Safety precautions 8 3 3 Maintenance routine 9 4 Calibration routine 9 4 1 General requirements 9 4 2 Preliminary arrangements 9 4 3 Safety precautions 10 4 4 Conditions 10 4 5 Procedure 10 5 Routine repairs 11 6 Storage 11 7 Transportation 11 8 Disposal 11 Appendix A Description of data exchange registers using the DiBUS pro...

Страница 3: ...a transfer to the external devices intended for visualization storage and alarm signalling Details on DiBUS protocol are presented at www doza ru Detectors are manufactured in modifications that differ by data exchange interface BDMN 200DD AJAH 418266 007 interface RS 422 BDMN 200PD AJAH 418266 007 01 interface RS 485 1 2 Technical characteristics 1 2 1 Energy range of measured neutron radiation 0...

Страница 4: ... operable within the limits stated in sections 1 2 3 1 2 10 during the whole life cycle under specified operation conditions 1 2 15 Degree of protection provided by casings against ingress of solid items and water IP65 1 2 16 Detectors are stable against electromagnetic interference of 2 grade according to the IEC 1000 4 8 93 IEC 1000 4 9 93 IEC 61000 4 2 95 IEC 61000 4 3 2006 IEC 61000 4 4 2004 I...

Страница 5: ...is not included in the delivery kit it is assembled by the customer using the cable plug from the delivery kit in accordance with the wiring diagram Appendix B 1 4 Design and operation 1 4 1 Detector module is comprised of scintillation detector based on LiF enriched in 6 Li isotop to 85 and ZnS Ag placed in the plexiglass for conversion of slow thermal neutrons into scintillations flashes of ligh...

Страница 6: ...eneration of the Alarm signal begins when the AEDR of neutron radiation exceeds corresponding threshold set by operator The Sliding algorithm allows smoothing of fluctuations of the count rate and prevents false alarms but at the same time the respond to dynamic changes of the radiation intensity is delayed to some extent The delay is determined by the algorithm parameters the interval time and th...

Страница 7: ...ner electronic circuit is galvanically isolated from the detector casing 1 5 Marking and sealing 1 5 1 Detector casings have a nameplate with the following information trademark and name of the manufacturer supplier reference designation of the detector works number of detector according to the manufacturer s system of numeration degree of ingress protection against solid items and water provided ...

Страница 8: ... 2 4 3 Correction of indications in the end of the range more than 0 4 of the upper limit of measurement range is performed by adjusting the dead time using the Tetra_Checker software Increasing dead time results in increasing indications at high count rates only 3 MAINTENANCE 3 1 General notes Maintenance is performed with the purpose of ensuring reliable long term operation of detectors 3 2 Safe...

Страница 9: ...al treatment with distilled water and drying with filter paper are not required Dry cleaning can be performed with any periodicity Detectors should be disconnected from the power supply before dry cleaning and or decontamination 3 3 2 Periodic maintenance Periodic maintenance consists of calibration of detectors 4 CALIBRATION ROUTINE 4 1 General requirements Calibration of detectors are performed ...

Страница 10: ...detector operability can be checked by presence of indication of a measurement result in corresponding window of the TETRA_Checker software 4 5 3 Determination of basic relative measurement error of AEDR of neutron radiation 4 5 3 1 Determine the basic relative measurement error by means of successive exposure of the detector to a set of AEDR of neutron radiation in the ranges the first value is s...

Страница 11: ... in the original manufacturer s package can be transported by all means of transport at any distance transportation by railway shall be carried out in clean boxcars when transported by air the boxes with detectors shall be placed in air tight heated compartment when transported by water and sea transport the boxes with detectors shall be placed in the hold 7 2 Arrangement and fastening of the boxe...

Страница 12: ...ctors accepted for operation after decontamination are subjects for repair or replacement in case of failure Detectors not suitable for operation with radioactive contamination levels below permissible limits should be dismantled to prevent further use and transferred to a special site for disposal of industrial wastes 8 4 Detectors with expired lifetime accepted for use after decontamination shal...

Страница 13: ... of selected measured quantity R Single 25 0 13 Measurement interval of selected measured quantity R DWORD 11 0 14 Uncertainty of the measurement result R BYTE 1 0 15 Selection of the measured quantity 1 R W BYTE 1 0 18 Status of the detector R WORD 5 0 19 Restart of measurement W BYTE 1 0 1c Detector code 2 R BYTE 1 0 1d Setting the net address of the block W DiBUS_address 33 0 1e The detector s ...

Страница 14: ...of the default measured quantity The measured quantity set as default is used by registers Instantaneous value 0 09 Combined package Measurement result of the selected measured quantity 0 0f Measurement result of the selected measured quantity 0 10 Measurement interval of the selected measured quantity 0 13 Uncertainty of the measurement result 0 14 А 3 Restart of measurement Register 0 19 is used...

Страница 15: ...ion of the register Selection of the measured quantity in the table A 1 calculated using data for the UniqSec th second InsPSS number of pulses acquired during UniqSec th second 4 bytes data type Single 25 UniqSec unique identifier changes every second 4 bytes data type DWORD 11 Examples of packages are presented in the Table A 4 Table A 4 Examples of packages Note Package Request for measurement ...

Страница 16: ...Note A1 Divider AJAH 418289 007 1 BL1 Photomultiplier tube FEU 35 1 1 A2 Amplifier AJAH 418288 007 04 1 A3 Voltage multiplier AJAH 418289 005 1 A4 Regulator AJAH 418283 012 1 BD1 Detector ZnS 1 30x4 mm HL1 LED L 34HD Kingbright 1 XP3 Plug RS10TV 1 XS1 Socket EHR 3 JST 1 XS2 Socket EHR 2 JST 1 ...

Страница 17: ...8266 007RE 17 Appendix C оbligatory OUTLINE DRAWING Figure С 1 Portable version of the detector with handle Figure С 2 Detector with light weight attachment fitting Detector module FVKM 305615 001 4 holes Ø6 5 ...

Страница 18: ...AJAH 418266 007RE 18 Figure С 3 Detector with wall mounting Wall mounting Neutron Moderator 4 holes Detector module ...

Страница 19: ...o the device nomenclature of parameters is defined by the device itself select one of the measured quantities in the device as a quantity requested by default indicate information about the operation of the device and measurement results on the monitor of personal computer indicate the device status information on the monitor of personal computer ATTENTION BUGS ARE POSSIBLE WHEN USING THE PROGRAM ...

Страница 20: ... the main window D 2 1 1 Parameters button The Parameters button is located in the upper right corner of the main program window Pressing the button Parameters displays a window shown in figure D 2 2 Figure D 2 2 Overview of the window Parameters ...

Страница 21: ...rrect port or release port by shutting down the concurrent program No reply this message appears when the device is not connected or the conversion mode of the matching device RS 232 to RS 485 RS 422 is incorrect Read error probably in the broadcast request mode several devices answer simultaneously it is recommended to switch to request mode with specific net address or working in the broadcast r...

Страница 22: ...he writing new parameter values will be displayed in the column Parameter point after a while ATTENTION DECIMAL POINT SEPARATOR BETWEEN INTEGER PART OF THE NUMBER AND THE FRACTION IS USALLY A DOT HOWEVER YOUR OPERATION SYSTEM CAN BE SET TO USE COMMA INSTEAD BE CAREFUL WHEN YOU INPUT NUMBERS Sets of values or ranges of values of parameters are indicated in the operational documentation for each dev...

Страница 23: ...values of parameters on the Device parameters panel are refreshed SET PARAMETERS F2 when the button F2 is pressed new values of the dynamic parameters are written into device At the same time parameter date of correction of parameters is automatically set to current date ...

Отзывы: