BUTTONS ON THE OPERATION PANEL
Clean/Ready
Button
Pump Button
Optic Button
Setting Button
Program Button
Power Button
Boost Button
Memory Button
It cleans out the air in the irrigation tube and creates consistent flow before and after surgery.
(Only irrigation pump will operate when you press Clean/Ready Button.
Handpiece will not operate)
Use for setting the irrigation volume, It is possible to set the level from 1 to 5.
This is Optic light selection button. This option only applied to piezo handpiece
with optic function.
I
t can control volume and reset unit’s settings.
- Volume Control: Manages volume from level 0 to 3
- Reset: Resets all of system settings.
It will return to primary menu once you press the button one more time.
Program selection is from 1 to 5.
Output power setting for piezo handpiece can be from level1 to 3.
Ultrasonic tip’s frequency setting from level 0 to 3 for adequate surgery environment.
Memorizes current program settings to selected number.
Level 1 (35 ㎖/min ±25%)
Level 2 (50 ㎖/min ±25%)
Level 3 (70㎖/min ±20%)
Level 4 (90 ㎖/min ±20%)
Level 5 (110 ㎖/min ±20%)
7
Содержание PIEZOART
Страница 1: ...USER MANUAL ...
Страница 23: ...MEMO 22 ...