DATA LOGGER SOUND LEVEL METER
26
Numerical Display:
4 digital Liquid Crystal Display per channel.
Measurement Range: -200°C ~ 1370°C
-328°F ~ 2498°F
Resolution
-200°C ~ 200°C
0.1°C; 200°C ~ 1370°C 1°C
-200°F ~ 200°F
0.1°F; else 1°F
Input Protection at
Thermocouple Input:
60V DC, or 24Vrms AC
Operation temperature: 0°C ~ 50°C (32°F ~ 122°F)
Operation humidity:
0 ~ 80% RH
Storage temperature: -10°C to 60°C (14°F ~ 140°F)
Storage humidity:
0 ~ 80% RH
Attitude:
up to 2000 meters
Electromagnetic
Compatibility:
Total accuracy = specified accuracy ± 2°C
(3.6°F)
Sample Rate:
3 seconds per period
Dimension:
184 x 64 x 30mm
Weight:
250g Approx.
Accessory:
K Type Bead Probex2, Battery, Carrying Case, I
Instruction Menu, Software program, RS-232
Connection Cable
Power requirement:
9 Volt Battery
Battery Life:
Approx. 100hrs with alkaline battery
AC Adapter:
9VDC ± 15% 100mA
Plug Diameter:
3.5mm x 135mm
Option:
AC Adapter
Temperature
Coefficient:
For ambient temperatures from 0°C ~ 18°C and
28°C ~ 50°C, for each °C ambient below 18°C
or above 28°C and the following tolerance into
the accuracy spec.
Accuracy: at ( 23 ± 5°C )
Range
Accuracy
-200°C ~ 200°C
±(0.2% r 1°C)
200°C ~ 400°C
±(0.5% r 1°C)
Содержание TC 309