23
22
Push (ENTER/MENUE) to confirm. On the second line of the display appears
Fd 0
.
Use the up and down keys (
) to select the requested range:
Fd 0
= below100,00°C
Fd 1
= above 100,00°C
Push (ENTER/MENUE) to confirm. On the display appears
00.00
. For changing the
value use the up key (
). For stepping to the next number use the down key (
). If the
requested value is complete then push (ENTER/MENUE) to confirm.
Note:
A break of the physical calibration cannot be done by the ESC-button. Switching off
the instrument can only do a break of the physical calibration.
Display-indication with active calibration code (
oP2
):
The
CAL
-segment and the small
2
on the left indicates to the user that
oP2
is
activated.
CAL
°C
CAL
°C
CAL
°C
1.3.5 Channel activation (Chnl)
only 2-channel instruments (P705-EX / P715-EX / P755-EX / P755-LOG-EX)/
Chnl = Channel
To activate or deactivate a measuring channel push (ENTER/MENUE). Use the up and down keys (
) to select
Chnl. Push again (ENTER/MENUE). On the left corner of the display appears a small 1, which indicates the selected
channel. To change the channel use the up and down keys (
). Push (ENTER/MENUE) to confirm. Use the up
and down keys (
) again to activate on or deactivate off the requested measuring channel and push (ENTER/
MENUE) to confirm. Push (ESC) to revert back to the measuring mode.
Note: As a minimum one channel is active!
MENUE
UniT
Prob
L in 2
CAL
Lo6
Chnl
nur / only
P710-EX/715-EX
P755-LOG-EX
nur / only
755-LOG-EX
nur / only
2-Kanal Geräte
nur / only
2-Kanal Geräte
1.3.6 Memory Setup Lo6
only for P755-LOG-EX
Push (ENTER/MENUE) and use the up and down keys (
) to select Lo6. Push (ENTER/MENUE) to
confirm. Use the up and down keys (
) again to start (off) or Stop (off) the logger. Push (ENTER/MENUE)
to confirm. Use the up and down keys (
) again to select between automatic storage (AUTo) or manually
operated storage (SPoT). Push (ENTER/MENUE) to confirm. Use the up and down keys (
) again to
select between to add on data (Add) and creating a new file (nLo6). Push (ENTER/MENUE) to confirm.
MENUE
UniT
Prob
L in 2
CAL
Lo6
Chnl
nur / only
P710-EX/715-EX
P755-LOG-EX
nur / only
755-LOG-EX
nur / only
2-Kanal Geräte
nur / only
2-Kanal Geräte
Lo6
on
oFF
1 S 1 Sekunde
5 S 5 Sekunde
10 S 10 Sekunde
20 S 20 Sekunde
30 S 30 Sekunde
1 M 1 Minute
2 M 2 Minute
5 M 5 Minute
10 M 10 Minute
20 M 20 Minute
Lo6
on
oFF
Add
nLo6
1 S
2 S
5 S...
nLo6
Add
1 S
2 S
5 S...
AUTo
SPoT
on
By selected automatic storage at the end you have to select the time interval:
E.g. of the instrument’s LCD by activated logging mode:
On the bottom you can see the percentage of the occupied memory (0..99%). If a
calibration option is activated the display is alternating between displaying memory
status and calibration information.
CAL
°C
1.4 Recalling the memory data (HOLD MAX MIN AVE)
After pushing first time the key (HOLD MAX MIN AVE) the actual values will be held on the display. Pushing
again the key (HOLD MAX MIN AVE), the saved maximum-, minimum and average values will be displayed.
Note:
During the recall of the memory data the extremes (MAX MIN) and the average value (AVE) will not
be calculated or carried on.
1.4.1 Clearing the memory (MAX MIN AVE)
Press (CLEAR) key once to erase the stored maximum, minimum and average from memory. On the
display appears
Clr.
– After erasing the memory the instrument automatically reverts back to measuring
mode indicating the actual measured value again.
1.5 Change Measuring rate (Normal-Mode / FAST-Mode)
The instrument has two different response times to select:
Normal-Mode:
one measurement per second
FAST-Mode:
approx. 4 measurements per second
After turning on the instrument it is in the Normal-Mode. After pressing once the (FAST / ) key the instrument
switches to the Fast-Mode.
E.g. Display with activated FAST-Mode:
The arrow on the left indicates that the instrument is
working in the FAST-Mode.
Note:
This setting is deactivated when the device is switched off.
°C
Содержание P700-EX
Страница 15: ...29 28 ...
Страница 16: ...31 30 ...