background image

 

 

Section 1. Unpacking and Installation 1 

Introduction .................................................................................................................. 1 
Unpacking .................................................................................................................... 1 
Cautions and Warnings ................................................................................................ 1 
Care & Cleaning ........................................................................................................... 1 
Installation .................................................................................................................... 2 
Battery Installation ........................................................................................................ 2 

Section 2.  Scale Operation ...................................................................... 3 

Display Pushbuttons ..................................................................................................... 3 
Weighing a Patient ....................................................................................................... 5 
Zero Operation ............................................................................................................. 5 
Reweigh Operation ....................................................................................................... 5 
Feed Function .............................................................................................................. 6 

Section 3.  Scale Calibration .................................................................... 7 

Multiple Point Calibration .............................................................................................. 7 
Calibration Troubleshooting ......................................................................................... 7 

Section 4.  Parameter Setup ..................................................................... 9 

Entering and Exiting Setup Mode ................................................................................. 9 
Setup Mode Navigation ................................................................................................ 9 
Calibration Weight Selection ........................................................................................ 9 
Calibration Mode Selections ......................................................................................... 9 
Start Up Zero .............................................................................................................. 10 
Automatic Shutoff Timer ............................................................................................. 10 
Raw Counts ................................................................................................................ 10 

Section 5. Specifications ........................................................................ 11 
Section 6.  Troubleshooting ................................................................... 12 

General Problem Resolution ...................................................................................... 12 
Error Messages .......................................................................................................... 13

Содержание DS4500

Страница 1: ...DS4500 Infant Lactation Scale Technical Manual MAN0348 Rev 1 Doran Scales Inc www doranscales com...

Страница 2: ...eweigh Operation 5 Feed Function 6 Section 3 Scale Calibration 7 Multiple Point Calibration 7 Calibration Troubleshooting 7 Section 4 Parameter Setup 9 Entering and Exiting Setup Mode 9 Setup Mode Nav...

Страница 3: ...ult in costly repairs that may not be covered by the warranty Please observe the following precautions to insure years of trouble free service from your new scale Do not use sharp objects to press any...

Страница 4: ...rd Automatic Shutoff Timer settings The standard Automatic Shutoff Timer setting is 10 seconds The batteries can be accessed on the bottom of the scale To access the battery compartment turn the scale...

Страница 5: ...not in use Zero When the zero button is pressed the scale will be reset to zero weight This feature can be used to zero out any weight not desired to be part of the patient weight Unit Press to cycle...

Страница 6: ...on the display This indicates an accurate weight has been achieved Low Batt Indicates when the batteries require replacement BEFORE FEED Indicates the next held weight recorded will be recorded for BE...

Страница 7: ...the scale platform 2 Press On Off The scale will turn on and display a zero weight If the item is removed a negative weight will be displayed Note Each time the scale is powered up the zero point is...

Страница 8: ...deactivated Instructions for Weighing 1 Turn on scale 2 Press Feed 3 Place infant on the scale 4 BEFORE FEED and HOLD annunciator is activated and scale beeps Display reads infant weight 5 Remove the...

Страница 9: ...ZERO and REWEIGH 5 Press and hold the FEED button until the display reads CALSP 6 Press UNIT and the display will read CAL 0 7 Remove all weight from the scale NOTE During calibration be sure vibrati...

Страница 10: ...tform and press FEED to perform the zero calibration again If ErrSp appears on the display during step 10 12 14 or 16 the span calibration is out of range Place the proper weight on the platform and p...

Страница 11: ...10 if the scale is in kg units To exit the setup mode press UNIT until the display returns to the weigh mode Setup Mode Navigation 1 Press UNIT to change the current parameter 2 Press FEED to change...

Страница 12: ...scale stays on after non use 10 10 seconds 30 30 Seconds 60 1 minute 120 2 minutes 180 5 minutes 300 10 minutes 600 15 minutes of Automatic shutoff timer disabled LB Units Enable Disable on Scale can...

Страница 13: ...5 W x 3 75 H Power 4 AA Alkaline Batteries One year battery life with 10 second Automatic Shutoff Timer setting Displayed Units lb lb oz kg Capacity 45 lb x 0 1 oz 45 x 0 005 lb 20 kg x 2 g Constructi...

Страница 14: ...cale Scale will not come to zero when the ZERO button is pressed Make sure that the scale is stabilizing After pressing the ZERO button the scale should zero as soon as it becomes stable Weight readin...

Страница 15: ...an calibration is out of range Error is displayed after a span calibration attempt Refer to the Calibration Troubleshooting section for additional information If this problem persists contact Doran Te...

Страница 16: ...Doran Scales Inc 883 Enterprise Court St Charles IL 60174 www doranscales com 630 879 1200...